5 - Front Panel Operation
Action
1.Displays the minimum current Displays the rms voltage Displays the input power Displays the maximum power Displays the minimum power
Display
<chan> <value> A MIN <chan> <value> A RMS <chan> <value> WATTS <chan> <value> W MAX <chan> <value> W MIN
Use‚ | and • | to scroll through the measurement selections |
7 - Setting the GPIB Address
Your electronic load is shipped with the GPIB address set to 5. This address can only be changed from the front panel using the Address menu located under the Address key
Set the GPIB address as follows:
| Action | Display |
1. | On the System keypad, press Address. | ADDRESS 5 |
2. | Enter the new address. For example, Press 7, Enter. | ADDRESS 7 |
8 - Storing and Recalling Instrument States
You can save up to 10 states (from location 0 to location 9) in
NOTE: The SAV0 state is the power up state.
Save an instrument state in location 0 as follows:
Action | Display |
1.Set the instrument to the state that you want to save.
2. Save this state to location 0. Press Shift, Save, 0, Enter.*SAV 0
Recall a saved state as follows:
| Action | Display |
1. | Recall the state saved in location 0 by pressing Recall, 0, Enter. | *RCL 0 |
Clear the
| Action | Display |
1. | On the System keypad, press the Recall key and scroll to the reset command. Then | *RST |
| press Enter. |
|
| This returns the unit to the |
|
2. | Save these settings to location 0. Press Shift, Save, 0, Enter. | *SAV 0 |
3. | Repeat step #2 for memory locations 7 through 9. | *SAV 7 |
|
| *SAV 8 |
|
| *SAV 9 |
70