B - Performance Test and Calibration

Performance Tests

IMON Zero Verification

This test verifies that IMON Zero is within calibration. IMON zero must be in calibration to perform the tests in this section. If the IMON value is not within calibration, go to the calibration section of this appendix and calibrate IMON zero, IPROG, current programming and readback and resistance programming and readback. See figure B-1 for IMON Zero test setup.

Action

1.Turn off load module and connect DVM to IMON as per figure B-1. Input terminals must be open ( no power source connected )

2.Turn on Load – press [ Recall } – scroll to *RST – press [Enter ]

3.Read IMON voltage from DVM.

4.If IMON in step 3 in not within specification then calibrate IMON zero, IPROG, Current mode and Resistance mode.

Normal Result

Voltage reading less than 0.166mV for models N3302A through N3305A Voltage reading less than 0.320mV for model N3306A

CC Mode Tests

These tests verify that the module operates in the CC mode and that IMON, current programming, and readback to the front panel display are within specification. Values read back over the GPIB are the same as those displayed on the front panel. To read back the current value via the GPIB, use MEAS:CURR:ACDC?

Current Programming and Measurement Accuracy Test

This test verifies that the current programming and measurement accuracy are within specification. If the test readings are out of tolerance the module may require calibration. If the test readings significantly disagree with the specified values or no readings can be obtained go to the Turn-On Checkout procedure to verify module operation.

Make a copy of the module test card for the model to be tested to record test values.

1.

Connect the Electronic Load, power source, DVM and current shunt as shown in figure B-2.

 

 

 

2.

Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ].

3.

Turn on the power source. Set the power source voltage and current to the values listed in the following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

5V

 

20V

 

5V

8V

5V

 

 

Power Source Current Setting

33A

 

12A

 

61A

61A

130A

 

4.

Checking high current range, Low input current. Press [ CURR ] [ 1 ] then [ Enter ]. Wait 10 seconds then record the

 

actual input current ( DVM reading / current shunt resistance ) and front panel current reading on the test card under

 

high current range low current.

 

 

 

 

 

 

 

 

 

 

 

5.

Checking high current range, maximum input current. Press [ CURR ] [ select current from following table] [Enter ].

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

 

N3305A

 

N3306A

 

 

 

 

30

10

 

60

 

 

60

 

120

 

 

 

Wait 30 seconds, then record the actual input current ( DVM reading / current shunt resistance ) and front panel current reading on the test card under high current range high current.

76

Page 73
Image 73
Agilent Technologies n3303a Performance Tests, Imon Zero Verification, CC Mode Tests, Performance Test and Calibration