Agilent Technologies n3300a, n3305a, n3304a, n3301a, n3303a, n3306a, n3302a manual Wait

Models: n3303a n3304a n3300a n3302a n3306a n3301a n3305a

1 102
Download 102 pages 27.56 Kb
Page 90
Image 90

 

 

 

Performance Test and Calibration - B

1110

!

 

 

1120

OUTPUT @Ld;"CAL:IPR:LEV P4"

 

1130

!

 

 

1140

PRINT TABXY(20,20),"Calibrating IMON & IPROG P4 - 30 SECOND WAIT"

1150

WAIT 30

 

 

1160

CLEAR SCREEN

 

 

1170

!

 

 

1180

INPUT "Measure voltage across current shunt and enter in volts",Ip4

1190

!

 

 

1200

INPUT "Measure Imon voltage and enter in volts",Vimonp4

1210

!

 

 

1220

INPUT "Measure Iprog voltage and enter in volts",Viprogp4

1230

!

 

 

1240

Ip4=Ip4/Rshunt

 

1250

!

 

 

1260

OUTPUT @Ld;"CAL:DATA ";Ip4,Vimonp4,Viprogp4

 

1270

!

 

 

1280

! *******

Calibrating Main Current DAC

********

1290

!

 

 

1300

PRINT TABXY(10,10),"Disconnect instruments from IMON and IPROG terminals"

1310

DISP "Press CONT when ready to calibrate main current ADC"

1320

PAUSE

 

 

1330

!

 

 

1340

CLEAR SCREEN

 

 

1350

OUTPUT @Ld;"CAL:LEV P1"

 

1360

PRINT TABXY(10,20),"Calibrating high current range P1 - 30 SECOND WAIT"

1370

WAIT 30

 

 

1380

CLEAR SCREEN

 

 

1390

!

 

 

1400

INPUT "Measure voltage across current shunt and enter in volts",Ip1

1410

!

 

 

1420

Ip1=Ip1/Rshunt

 

1430

!

 

 

1440

OUTPUT @Ld;"CAL:DATA ";Ip1￿￿1450 !￿￿￿1460

OUTPUT @Ld;"CAL:LEV P2"

1470

!

 

 

1480

PRINT TABXY(10,10),"Calibrating high current range P2 - 30 SECOND WAIT"

1490

WAIT 30

 

 

1500

CLEAR SCREEN

 

 

1510

!

 

 

1520

INPUT "Measure voltage across current shunt and enter in volts",Ip2

1530

!

 

 

1540

Ip2=Ip2/Rshunt

 

1550

!

 

 

1560

OUTPUT @Ld;"CAL:DATA ";Ip2

 

1570

!

 

 

1580

OUTPUT @Ld;"INP:STAT OFF"

 

1590

OUTPUT @Ld;"CURR:RANG MIN"

 

1600

OUTPUT @Ld;"SENS:CURR:RANG MIN"

 

1610

!

 

 

1620

CLEAR SCREEN

 

 

1630

!

 

 

1640

INPUT "Enter value of low range current shunt in ohms",Rshuntl

1650

DISP "You have entered ";Rshuntl;" ohms"

 

1660

INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$

1670

IF UPC$(Y$)="Y" OR Y$="" THEN 1690

 

1680

IF UPC$(Y$)="N" OR Y$="n" THEN 1620

 

93

Page 90
Image 90
Agilent Technologies n3300a, n3305a, n3304a, n3301a, n3303a, n3306a, n3302a manual Wait