B - Performance Test and Calibration

Agilent N3303A Verification Test Record

Test Description

Minimum

Results

Maximum

 

 

Specification

 

Specification

 

Constant Current Mode Tests

 

 

10 Ampere Range Programming and Readback

0.9915 A

_______A

1.041 A

Low Current ( 1 A )

 

Front Panel Display

 

Aout – 5.5mA

_______A

Aout + 5.5mA

High Current ( 10 A )

 

9.9825 A

_______A

10.0175 A

Front Panel Display

 

Aout – 10mA

_______A

Aout + 10mA

1 Ampere Range Programming and Readback

0.0959 A

_______A

0.1041 A

Low Current ( 0.1 A )

 

Front Panel Display

 

Aout – 2.55 mA

_______A

Aout + 2.55mA

High Current ( 1 A )

 

0.995 A

_______A

1.005 A

Front panel Display

 

Aout – 3mA

_______A

Aout + 3mA

 

 

 

 

 

 

Current Regulation test

 

 

1.0 Amp Regulation at input voltage delta 4V to 240V

 

______mA

Less then 8mA

 

 

 

 

 

 

Voltage Mode Tests

 

 

240 Volt Range Programming and Readback

2.957 V

_______V

3.043 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 21mV

_______V

Vout + 21mV

High Voltage ( 240 V )

 

239.720 V

_______V

240.280 V

Front Panel Display

 

Vout – 140mV

_______V

Vout + 140mV

24 Volt Range Programming and Readback

2.987 V

_______V

3.013 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 11mV

_______V

Vout + 11mV

High Voltage ( 24 V )

 

23.966 V

_______V

24.034 V

Front Panel Display

 

Vout – 22mV

_______V

Vout + 22mV

 

 

 

 

 

 

Voltage Mode

Regulation Test

 

 

25V regulation with input current delta 0.4A to 10A

 

______mV

Less then 10mV

 

 

 

 

 

 

Resistance

Mode Tests

 

 

Resistance Range 1

 

 

______ Ω

49.54 Ω

High Resistance Point

48 Ω

46.46 Ω

Low Resistance Point

1.2 Ω

1.088 Ω

______ Ω

1.312 Ω

Resistance Range 2

 

41.30 Ω

______Ω

46.70 Ω

Low Resistance Point

44 Ω

High Resistance Point

480 Ω

383.50 Ω

______Ω

576.5 Ω

Resistance Range 3

 

325 Ω

______Ω

598.4 Ω

Low Resistance Point

440 Ω

High Resistance Point

4800 Ω

2640 Ω

______Ω

38,400 Ω

Resistance Range 4

 

1496 Ω

______Ω

39,600 Ω

Low Resistance Point

4400 Ω

High Resistance Point

12,000 Ω

4080 Ω

______Ω

108,000 Ω

 

 

 

 

 

 

Transient Generator Tests

 

 

Frequency Range Tests (1000 Hz test)

995 Hz.

_____ Hz

1005 Hz

Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz

49%

______ %

51%

Pulse Width Test (100∝S wide; 10 A pulse)

99∝s

______s

101∝s

 

 

 

 

 

84

Page 81
Image 81