A

Specifications

Table A-1 lists the specifications for the different load models. Specifications indicate warranted performance in the 25°C ±5°C region of the operating temperature range. Specifications apply to normal and transient modes unless otherwise noted.

 

Table A-1. Specifications

 

 

 

 

N3302A

N3303A

N3304A

 

N3305A

N3306A

Input Ratings

 

 

 

 

 

 

Current

0 - 30 A

0 - 10 A

0 - 60 A

 

0 - 60 A

0 - 120 A

Voltage

0 - 60 V

0 - 240 V

0 - 60 V

 

0 - 150V

0 - 60V

Maximum Power @ 40°C 1

150 W

250 W

300 W

 

500 W

600 W

Input Characteristic

OPERATING CONTOUR

 

 

DERATED CURRENT DETAIL

 

 

 

 

Voltage

 

 

Voltage

 

 

full

 

 

 

 

 

 

scale

 

 

 

 

 

 

 

Max Power Contour

 

all specifications apply

 

 

 

 

 

 

 

 

 

 

 

3

 

 

 

 

 

 

2

slew rate limitations apply

 

 

 

 

 

(see Table A-2)

 

 

 

 

 

1

 

 

 

 

 

 

 

 

0

 

Current

0

 

Current

 

 

full

 

 

 

full

 

 

scale

 

 

 

scale

Specified Current @

 

 

 

 

 

Low Voltage Operation

 

 

 

 

 

2.0 V

30 A

10 A

60 A

60 A

120 A

1.5 V

22.5 A

7.5 A

45 A

45 A

90 A

1.0 V

15 A

5 A

30 A

30 A

60 A

0.5 V

7.5 A

2.5 A

15 A

15 A

30 A

0 V

0 A

0 A

0 A

0 A

0 A

1Maximum continuous power available is derated linearly from 100% of maximum at 40°C, to 75% of maximum at 55°C.

Typical Minimum Operating Voltage @ Full Scale Current

 

N3302A

N3303A

N3304A

N3305A

N3306A

Table A-1 states that maximum

1.2 V

1.2 V

1.2 V

1.4 V

1.4 V

current is available down to 2

 

 

 

 

 

volts. Typically, under normal

 

 

 

 

 

operating conditions, the load

 

 

 

 

 

can sink the maximum current

 

 

 

 

 

down to the following voltages:

 

 

 

 

 

71

Page 68
Image 68
Agilent Technologies n3302a, n3305a, n3304a, n3301a, n3303a, n3306a, n3300a manual Specifications