Description of Calibration Sequence

The Calibration Sequence sets the scale to an accurate reference point for weighing. This process involves the use of a Field Standard Weight Set (31.5-pounds) for calibration in pounds, (18.5-kilograms) for Metric. Once calibration has been successfully completed, the scanner/scale uses the certified weight as a reference for subsequent weighing activities.

These verification procedures follow the U.S. National Institute of Standards and Technology 44 Handbook guidelines for bench/counter scale installations.

If any of these tests fail to meet the required weight indications, you must calibrate the scanner/scale. Refer to the calibration procedures in this section for the proper procedures.

You may be required by state and/or local regulations to have procedures other than these performed by a certified technician or verification offi- cial.

Access to the calibration switch should be restricted with a paper or a wire and lead seal after the calibration has been performed if required by your local regulatory agency.

The Calibration Sequence must be performed without removing the scanner/ scale from its installed position.

NOTE

The following tools and supplies will be required to perform the calibration process:

31.5 pound Field Standard Weight Set1. (English calibration only).18.5 kilogram Field Standard Weight Set2. (Metric calibration only).Lead/Wire or Paper Seal (as required by law).

1.NOTE: Throughout the calibration procedures, specified weights may be achieved by using a com- bination of weights from this set. eg. 20 lbs may be made up of one 10.00 lb. weight and two 5.00 lb. weights.

2.NOTE: Throughout the calibration procedures, specific weights may be achieved by using a com- bination of weights from this set. eg. 10 kgs. may be made up of one 5.00 kg. and five 1.00 kg. weights.