Fluke 8842a instruction manual 20Ω Ranges, Reduction of Zero-Input Error

Models: 8842a

1 269
Download 269 pages 45.95 Kb
Page 105
Image 105

8842A

Instruction Manual

The zero-input error is quickly reduced when the input is increased. The rms converter error (a dc error) and the internally generated noise (a random ac error) are both uncorrelated with the input signal. Therefore, when a signal is applied, the resulting reading is not the simple addition of the signal and the zero-input error, but the square root of the sum of their squares. This reduces the effect of the error, as shown in the example in Figure 4-9.

f4-09.wmf

Figure 4-9. Reduction of Zero-Input Error

As long as the 8842A reading is 1,000 counts or more, readings will still be within specified accuracy.

4-23. MAKING ACCURATE MEASUREMENTS ON THE 20 mV

AND 20Ω RANGES

NOTE

When making low-level (uV) measurements after a large step signal has been applied to the inputs, allow sufficient time for thermal emfs and other system-related sources of error to settle before taking readings.

The 20 mV dc and 20Ω ranges are the 8842A’s most sensitive ranges. For that reason, they are also the most susceptible to error from electrical noise, thermal voltages, and (for resistance measurements) test lead resistance. You can minimize these sources of error by using good measurement practices.

The most common source of error is electrical noise. Typical sources of noise include electrostatic noise, inductive pickup noise, radio frequency noise, power line noise and noise generated by ground loop currents. Noise pickup can be minimized by properly shielding the test leads between the 8842A and the signal source.

For voltage measurements in most system applications, where common-mode voltages are typically present, connect the test lead shielding to the 8842A INPUT LO terminal as shown in Figure 4-10. This configuration minimizes the error caused by current that would flow in the leads due to common-mode voltages between the measurement and stimulus points and instrument ground. The 8842A’s INPUT LO terminal is internally connected to the instrument’s internal guard, which provides a shield between the instrument’s ground and its sensitive analog circuits. The 8842A’s analog circuits are isolated from its digital circuits by an electrostatically shielded transformer, whose shield is also connected to the guard.

4-14

Page 105
Image 105
Fluke 8842a instruction manual 20Ω Ranges, Reduction of Zero-Input Error