Fluke 8842a instruction manual Hours 23±1C

Models: 8842a

1 269
Download 269 pages 45.95 Kb
Page 22
Image 22

8842A

Instruction Manual

Accuracy

NORMAL (S) READING RATE

...............±(% of Reading + Number of Counts)1.

RANGE

20Ω3

200Ω3

2 kΩ

20 kΩ

200 kΩ

2000 kΩ

20 MΩ

24 HOURS 23±1°C

0.007 + 304

0.0040 + 35

0.0025 + 2

0.0025 + 2

0.0025 + 2

0.023 + 3

0.023 + 3

90 DAY 23±5°C

0.009 + 404

0.007 + 45

0.005 + 3

0.005 + 3

0.006 + 3

0.025 + 3

0.040 + 4

1 YEAR 23±5°C

0.012 + 404

0.010 + 45

0.008 + 3

0.008 + 3

0.010 + 3

0.027 + 3

0.042 + 4

2 YEARS 23±1°C

0.015 + 404

0.012 + 45

0.010 + 3

0.010 + 3

0.012 + 3

0.030 + 3

0.050 + 4

1.Within one hour of ohms zero, using offset control.

2.Relative to calibration standards.

3.Applies to 4-wire ohms only.

4.When offset control is not used the number of counts are 50, 70, 90 and 90 for 24 hours,90 day, 1 year, and 2 year respectively.

5.When offset control is not used the number of counts are 5, 7, 9 and 9 for 24 hours, 90 day, 1 year, and 2 year respectively.

MEDIUM AND FAST READING RATES

........In medium rate, add 2 counts to the number of counts for

 

the 200Ω through 200 kΩ ranges, 3 counts for the 2000

 

kΩ and 20 MΩ ranges, and 20 counts for the 20Ω range.

 

In fast reading rate, use 3 (4½ digit mode) counts for the

 

number of counts for the 200Ω range, 20 (4½ digit mode)

 

counts for the 20Ω range and 2 (4½ digit mode) counts for

 

all other ranges.

Operating Characteristics

 

TEMPERATURE COEFFICIENT

Less than 0.1 x accuracy specification per °C from 0°C to 18°C

 

and 28°C to 50°C.

MEASUREMENT CONFIGURATION

2-wire or 4-wire in all ranges accept 20Ω range. Only 4-wire

 

configuration is allowed in the 20Ω range.

OPEN CIRCUIT VOLTAGE

Less than 6.5V on the 20Ω through the 200 kΩ ranges. Less

 

than 13V on the 2000 kΩ and 20 MΩ ranges.

INPUT PROTECTION

To 300V rms.

1-8

Page 22
Image 22
Fluke 8842a instruction manual Hours 23±1C