8842A

Instruction Manual

CURRENT

Input Characteristics

RANGE

200 mA 2

2000 mA

FULL SCALE 5½

DIGITS

199.999 mA

1999.99 mA

 

RESOLUTION

 

 

 

5½ DIGITS

 

1

 

4½ DIGITS

1 μA

 

10 μA

10 μA

 

100 μA

 

 

 

1.4½ digits at the fastest reading rate.

2.The 200mA range is available for dc current only.

DC Accuracy

NORMAL (S) READING RATE

±(% of reading + number of counts).

RANGE

200 mA

2000 mA

1A

>1A

90DAYS 23±5°C

0.04+ 40

0.04+ 4 0.1 + 4

1YEAR 23±5°C

0.05+ 40

0.05+ 4

0.1+ 4

2YEARS 23±5°C

0.08+ 40

0.08+4

0.15+4

MEDIUM AND FAST READING RATES In medium reading rate, add 2 counts (20 counts on 200 mA range) to number of counts. In fast reading rate, use 2 (4½ digit mode) counts (20 counts on 200 mA range) for number of counts.

AC Accuracy (Option –09)

 

 

NORMAL (S) READING RATE

...............±(% of Reading + Number of Counts).

 

 

23±5°C, for sinewave inputs 10,000 counts1.

 

 

 

 

 

FREQUENCY IN HERTZ

 

 

 

ONE YEAR

TWO YEAR

*Typically 20 kHz

20-45

2.0+ 200

3.0+ 300

45-100

0.5+ 200

0.7+ 300

100-5K*

0.4+ 200

0.6+ 300

1.For sinewave inputs between 1,000 and 10,000 counts, add to number of counts 100 counts for frequencies 20 Hz to 5 kHz (typically 20 kHz).

1-6

Page 20
Image 20
Fluke 8842a instruction manual Current, Full Scale 5½ Digits