Fluke 8842a instruction manual General, Power

Models: 8842a

1 269
Download 269 pages 45.95 Kb
Page 25
Image 25

 

Introduction and Specifications

 

 

SPECIFICATIONS

1

GENERAL

 

 

COMMON MODE VOLTAGE

1000V dc or peak ac, or 700V rms ac from any input to earth.

 

TEMPREATURE RANGE

0 to 50°C operating, -40 to 70°C storage.

 

HUMIDITY RANGE

80% RH from 0 to 35°C, 70% to 50°C.

 

WARMUP TIME

1 hour to rated specifications.

 

POWER

100, 120, 220, or 240V ac ±10% (250V ac maximum), switch

 

 

selectable at rear panel. 50, 60, or 400 Hz, automatically

 

 

sensed at power-up. 20 VA maximum.

 

VIBRATION

Meets requirements of MIL-T- 28800C for Type III, Class 3,

 

 

Style E equipment.

 

PROTECTION

ANSI C39.5 AND IEC 348, Class I.

 

SIZE

8.9 cm high, 21.6 cm wide, 37.1 cm deep(3.47 in high, 8.5 in

 

 

wide, 14.6 in deep).

 

WEIGHT

Net, 3.4 kg (7.5 lb); shipping, 5.0 kg (11 lb).

 

INCLUDED

Line cord, test leads, Instruction/Service Manual, IEEE-488

 

 

Quick Reference Guide, (Option –05 only), and instrument

 

 

performance record.

 

IEEE-488 INTERFACE FUNTION

Option allows complete control and data output capability, and

 

 

supports the following interface funtion subsets: SH1,AH1, T5,

 

 

L4, SR1, RL1, DC1, DT1, E1, PP0, AND C0.

 

ELECTROMAGNETIC COMPATIBILITYSpecifications apply when used in an environment with fields strengths 1 V/m, (0.8 V/m for DC Current.) For fields strengths up to 3 V/m, multiply floor adder by 12 for VDC and Resistance and 200 for DC current. VAC and AC Current have no adders up to 3 V/m.

f1-01.wmf

Figure 1-1. External Dimensions

1-11