Fluke 8842a instruction manual Device-Dependent Command Set

Models: 8842a

1 269
Download 269 pages 45.95 Kb
Page 49
Image 49

8842A

Instruction Manual

FUNCTON COMMANDS

F1

VDC (Default)

F2

VAC

F3

2 WIRE kΩ

F4

4 WIRE kΩ

F5

MA DC

F6

MA AC

RANGE COMMANDS

R0

Autorange

R2

2V, 2 kΩ

R3

20V, 20 kΩ

R5

1000V dc, 700V ac, 2 MΩ, 2000 mA

R6

20 mΩ

R8

20mV, 20Ω

READING RATE COMMANDS

S0 Slow (Default)

S1 Medium

S2 Fast

TRIGGER MODE COMMANDS

COMMAND

TRIGGER

REAR PANEL

SETTLING

 

MODE

TRIGGER

DELAY

TO (Default)

Internal

Disabled

-

T1

External

Enabled

On

T2

External

Disabled

On

T3

External

Enabled

Off

T4

External

Disabled

Off

Note: Delay is enabled by entering EX TRIG mode while in local.

OFFSET COMMANDS

B0

B1

DISPLAY COMMANDS

D0 Offset Off (Default)

D1 Offset On

SUFFIX COMMANDS

Y0 Normal Display (Default)

Y1 Blank Display

TERMINATOR COMMANDS

W0 Enable CR LF EOI (Default)

W1 Enable CR LF Only

W2 Enable CR EOI Only

W3 Enable CR Only

W4 Enable LF EOI Only

W5 Enable LF Only

W7 Disable All Output Terminators

CLEAR COMMANDS

*Device Clear (Resets 8842A to default conditions)

X0 Clear Error Register

SINGLE-TRIGGER COMMAND

?Trigger Measurement

GET COMMANDS

G0 Get Instrument Configuration (F,R,S, and T)

G1 Get SRQ Mask

G2 Get Calibration Input Prompt

G3 Get User-Defined Message

G4 Get Calibration Status

G5 Get IAB Status (input F/R, Autorange On/Off, Offset On/Off)

G6 Get YW Status (Suffix

Enabled/Disabled, Terminator

Selection)

G7 Get Error Status

G8 Get Instrument Identification Note: G2 valid only in calibration mode.

PUT COMMANDS

P0 Put Instrument Configuration (F,R,S, and T)

P1 Put SRQ Mask

P2 Put Variable Calibration Value

P3 Put User-Defined Message

Note: P2 and P3 valid only in calibration mode.

PUT COMMAND FORMAT

N<value> P0

N <value> P1

N <value> P2

P3 < 16 ASCII characters>

Figure 3-5. Device-Dependent Command Set

3-8

Page 49