8842A
Instruction Manual
FUNCTON COMMANDS
F1 | VDC (Default) |
F2 | VAC |
F3 | 2 WIRE kΩ |
F4 | 4 WIRE kΩ |
F5 | MA DC |
F6 | MA AC |
RANGE COMMANDS | |
R0 | Autorange |
R2 | 2V, 2 kΩ |
R3 | 20V, 20 kΩ |
R5 | 1000V dc, 700V ac, 2 MΩ, 2000 mA |
R6 | 20 mΩ |
R8 | 20mV, 20Ω |
READING RATE COMMANDS
S0 Slow (Default)
S1 Medium
S2 Fast
TRIGGER MODE COMMANDS
COMMAND | TRIGGER | REAR PANEL | SETTLING | |
| MODE | TRIGGER | DELAY | |
TO (Default) | Internal | Disabled | - | |
T1 | External | Enabled | On | |
T2 | External | Disabled | On | |
T3 | External | Enabled | Off | |
T4 | External | Disabled | Off | |
Note: Delay is enabled by entering EX TRIG mode while in local.
OFFSET COMMANDS
B0
B1
DISPLAY COMMANDS
D0 Offset Off (Default)
D1 Offset On
SUFFIX COMMANDS
Y0 Normal Display (Default)
Y1 Blank Display
TERMINATOR COMMANDS
W0 Enable CR LF EOI (Default)
W1 Enable CR LF Only
W2 Enable CR EOI Only
W3 Enable CR Only
W4 Enable LF EOI Only
W5 Enable LF Only
W7 Disable All Output Terminators
CLEAR COMMANDS
*Device Clear (Resets 8842A to default conditions)
X0 Clear Error Register
?Trigger Measurement
GET COMMANDS
G0 Get Instrument Configuration (F,R,S, and T)
G1 Get SRQ Mask
G2 Get Calibration Input Prompt
G3 Get
G4 Get Calibration Status
G5 Get IAB Status (input F/R, Autorange On/Off, Offset On/Off)
G6 Get YW Status (Suffix
Enabled/Disabled, Terminator
Selection)
G7 Get Error Status
G8 Get Instrument Identification Note: G2 valid only in calibration mode.
PUT COMMANDS
P0 Put Instrument Configuration (F,R,S, and T)
P1 Put SRQ Mask
P2 Put Variable Calibration Value
P3 Put
Note: P2 and P3 valid only in calibration mode.
PUT COMMAND FORMAT
N<value> P0
N <value> P1
N <value> P2
P3 < 16 ASCII characters>