Fluke 8842a instruction manual Erasing Calibration Memory, Tolerance Check

Models: 8842a

1 269
Download 269 pages 45.95 Kb
Page 155
Image 155

8842A

Instruction Manual

1.If you just completed the Offset and Gain or High-Frequency AC Calibration for an entire function (not just one range), the range buttons can be used to change ranges in order to verify all ranges were calibrated correctly.

2.If you just calibrated an individually selected range, pressing another range button begins the Offset and Gain Calibration procedure for the new range.

3.You can use the RATE button to verify the calibration at other reading rates.

CAUTION

It is still possible to erase the calibration memory while the 8842A is taking verification readings.

6-19. ERASING CALIBRATION MEMORY

The 8842A allows you to erase some or all of the correction constants stored in calibration memory (U220). It is recommended that you erase the entire calibration memory before beginning calibration if the calibration memory is replaced or accidentally altered. The capability of erasing particular parts of the memory is mainly intended as a troubleshooting aid to the technician.

CAUTION

Once the calibration memory is erased, the 8842A must be recalibrated.

To erase all or part of the calibration memory, proceed as follows:

1.Press the front panel ERASE button. The display should show the erase prompt "cl" (for "clear"). If you do not press another button within 1-1/2 seconds, the 8842A returns to its previous state.

2.To complete an erasure, press one of the following buttons within 1-1/2 seconds of pressing the ERASE button:

a.STORE -- Erases the entire memory.

b.A/D -- Erases the A/D Calibration constants

c.Any function button -- Erases the Offset and Gain Calibration constants for all ranges of that function.

d.HF AC -- Erases the High-Frequency AC Calibration constants.

3.After an erasure is finished (a complete erasure takes about 3 seconds), the 8842A returns to one of the following states:

a.After complete erasure: Begins A/D Calibration.

b.After A/D erasure: Begins A/D Calibration.

c.After Offset and Gain erasure: Begins Offset and Gain Calibration for erased function.

d.After High-Frequency AC erasure: Begins High-Frequency AC Calibration.

6-20. TOLERANCE CHECK

The 8842A automatically checks that the reference input is near the value prompted. This minimizes common errors such as applying a reference source with the wrong sign. If the reference input exceeds the tolerances shown in Table 6-13, the 8842A displays ERROR 41.

6-18