Fluke 8842a instruction manual 24. P1 Put SRQ Mask, 25. P2 Put Calibration Value

Models: 8842a

1 269
Download 269 pages 45.95 Kb
Page 57
Image 57

8842A

Instruction Manual

"N3120 P0"

Identical to F3 R1 S2 T0. Selects 2 WIRE kΩ function, 200Ω range,

 

fast sample rate, continuous trigger.

The P0 command allows broadside loading of the Function, Range, Reading Rate, and Trigger Mode commands (F, R, S, and T). The codes for these commands are listed in Figure 3-5.

A numeric entry for P0 must be within +1000 and +6824. Each of the four digits must not exceed its maximum allowed value (6, 8, 2, and 4, respectively) or an error message will occur and the instrument configuration will remain unchanged. The entry may be expressed as an integer, real number, or real number with exponent, as described under the N command. Any fractional part is ignored.

Example Explanation

"N3112 P0"

Sets the 8842A to F3, R1, S1, and T2.

3-24. P1 (Put SRQ Mask)

FormatExplanation

N<SRQ mask>P1

Where <SRQ mask> is a two-digit integer from 00 to 63.

The P1 command is used to program the 8842A to make service requests on user- specified conditions. The two-digit code for the SRQ mask is interpreted in Table 3-1 under the G1 command. For more about the SRQ mask, see paragraph 3-51.

Numeric entries for the P1 command must be between 0 and +63 (inclusive), or an error will occur and the SRQ mask will remain unchanged. The entry may be expressed as an integer, real number, or real number with exponent, as described under the N command. Any fractional part is ignored.

Example Explanation

"N0.17E+2 P1" Sets SRQ mask to 17. Enables SRQ on data available or overrange.

"N1 P1"

Sets SRQ mask to 01. (A leading zero is assumed.) Enables SRQ on

 

overrange.

3-25. P2 (Put Calibration Value)

FormatExplanation

N<value>P2

Where <value> can be an integer, real number, or real number with

 

exponent, as described under the N command.

Example

Explanation

"N1 P2"

If the 8842A is in VDC, the next calibration input expected is

 

1.00000V dc.

The P2 command is used to enter variable input calibration values just like the front panel VAR IN button. To accept the P2 command, the 8842A must be in the calibration mode (enabled by pressing the front panel CAL ENABLE switch). Otherwise, the P2 command will generate an error message.

The variable input is a measurement value that is to be used as the calibration value for the next calibration step. Its format is the same as a measurement value. But since it is coming from the controller, the value can be specified using any valid format (signed interger, real number, or real number with exponent). For example, if the 8842A prompts for an input value of 100Ω for the next calibration step, but the available source is

98.97Ω, the variable input can be specified as "N+9.897E+1", "N0.9897E2", N9897E-2",

3-16

Page 57
Image 57
Fluke 8842a instruction manual 24. P1 Put SRQ Mask, 25. P2 Put Calibration Value