Fluke 8842a instruction manual 15. G3 Get User-Defined Message, 16. G4 Get Calibration Status

Models: 8842a

1 269
Download 269 pages 45.95 Kb
Page 54
Image 54

Remote Programming 3

DEVICE-DEPENDENT COMMAND SET

3-15. G3 (Get User-Defined Message)

The G3 command loads the output buffer with the user-defined message stored in calibration memory during the calibration procedure. The message consists of 16 ASCII characters, as shown in Figure 3-6.

The message is stored in calibration memory during calibration using the P3 command. If fewer than 16 characters have been stored, the remaining characters returned are spaces. If no message has ever been stored, a string of 16 null characters (hex 00) will be returned. Some example output strings follow.

Example

Explanation

FL8842A.12-17-83 CR LF

Identifies instrument and gives cal date.

01-25-84 CR LF

Gives cal date. The last eight characters are

 

blank.

3-16. G4 (Get Calibration Status)

The G4 command is used when calibrating the 8842A under remote control. The command loads the output buffer with the instrument’s calibration status in the format shown in Figure 3-6. The status is represented by a four-digit integer which is interpreted in Table 3-1.

The first two digits are always 1 and 0. The third digit indicates whether or not the calibration verification mode is enabled. (This mode is enabled only when the calibration mode is enabled.) The fourth digit indicates whether or not the calibration mode is enabled, and if so, which part of the calibration procedure the 8842A is in. Example output strings follow.

Example Explanation

1000 CR LF 1: Leading 1

0: Leading 0

0:Not in cal verification

0:Cal mode disabled

1001 CR LF 1: Leading 1

0: Leading 0

0:Not in cal verification

1:Cal mode enabled; A/D cal selected

3-13

Page 54
Image 54
Fluke 8842a instruction manual 15. G3 Get User-Defined Message, 16. G4 Get Calibration Status