8842A

Instruction Manual

f6-01.wmf

Figure 6-1. DC Calibration Connections

STEP

RANGE

INPUT

(V dc)

Table 6-2. DC Voltage Test

Displayed Reading

 

SLOW

 

MEDIUM

 

 

FAST

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MIN

 

MAX

 

MIN

 

MAX

MIN

 

MAX

 

 

 

 

 

 

 

 

 

 

 

 

A1

20 mV

0V

-0.0030

+0.0030

-0.0050

+0.0050

-0.030

 

 

(short)

 

 

 

 

 

B1

200 mV

0V

 

 

 

 

 

-00.003

+00.003

-00.005

+00.005

-00.03

 

 

(short)

 

 

 

 

 

C1

2V, 20V,

0V

-2 counts

+2 counts

-4 counts

+ 4 counts

-3 counts

 

200V,

(short)

 

 

 

 

 

 

1000V

 

 

 

 

 

 

D

20 mV

10 mV

 

 

 

 

 

+9.9963

+10.0037

+9.9943

+10.0057

+9.970

E

200 mV

100 mV

 

 

 

 

 

+99.993

+100.007

+99.991

+100.009

+99.97

F

2V

1V

 

 

 

 

 

+.99996

+1.00004

+.99994

+1.00006

+.9997

G

20V

10V

 

 

 

 

 

+9.9995

+10.0005

+9.9993

+10.0007

+9.997

H

200V

100V

 

 

 

 

 

+99.995

+100.005

+99.993

+100.007

+99.97

I

1000V

1000V

 

 

 

 

 

+999.95

+1000.05

+999.93

+1000.07

+999.7

 

 

 

 

 

 

 

 

1. Relative to high-quality short stored using OFFSET feature.

+0.030

+00.03

+3 counts

+10.030

+100.03

+1.0003

+10.003

+100.03

+1000.3

6-6

Page 143
Image 143
Fluke 8842a instruction manual Step Range Input, Slow Medium Fast MIN MAX