8842A
Instruction Manual
AC Calibrator | Fluke 5700A and Fluke 5725A |
|
|
Minimum Required Accuracy (By Range)
Frequency Range
1, 10, 100 mV1
1, 10, 100V2
1000V2
20 Hz – 30 Hz | .1 + .005 |
| .1 + .005 | .12 | + .005 |
30 Hz – 20 kHz | .02 + 10 |
| 0.2 + .002 | .04 | + .004 |
20 kHz – 50 kHz |
|
| .05 + .005 |
| + .005 |
.05 + 20 |
| .08 | |||
50 kHz – 100 kHz |
|
| .05 + .005 |
| + .01 |
.05 + 20 |
| .1 | |||
1. ±(% of setting + μV) 2. |
|
|
| ||
AC Current Source |
|
| Fluke 5700A |
|
|
|
|
|
|
|
|
Frequency Range
Minimum Required Accuracy (All Ranges)
30 Hz – 1 kHz 1kHz – 5 kHz
±.07% + 1 mA)
±(.07% + 1 mA) X frequency in kHz
Shorting Bar
Optional Test Equipment
Resistance <1.5 mΩ | Pomona |
Construction Soldered (not |
|
rivetted) |
|
9010A, 9005A or
6-2. PERFORMANCE TEST
This test compares the performance of the 8842A with the specifications given in Section
1.The test is recommended as an acceptance test when the instrument is first received, and as a verification test after performing the calibration procedure. If the instrument does not meet the performance test, calibration or repair is needed.
To ensure optimum performance, the test must be performed at an ambient temperature of 18°C to 28°C, with a relative humidity of less than 75%. Also, the 8842A should be allowed to warm up for one hour prior to beginning any test other than the
6-3. Diagnostic Self-Tests
The diagnostic
NOTE
The inputs must be left
If the FRONT/REAR switch is in the REAR position, the 8842A skips tests 3 and 4. Also, if Option