8842A Instruction Manual

Operating Characteristics

 

TEMPERATURE COEFFICIENT

...........<(0.0006% of reading + 0.3 Count) per °C from 0°C to 18°C

 

and 28°C to 50°C.

MAXIMUM INPUT

1000V dc or peak ac on any range.

NOISE REJECTION

Automatically optimized at power-up for 50, 60, or 400 Hz.

RATE

READINGS/ SECONDî

FILTER

NMRRï

PEAK NM

SIGNAL

CMRRð

S

2.5

Analog & Digital

>98 dB

20V or 2x FS

>140 dB

M￿

 

Digital

>45 dB

1x FS

>100 dB

20

F

 

None

 

1x FS

>60 dB

100

_

 

 

 

 

 

 

1.Reading rate with internal trigger and 60 Hz power line frequency. See “reading rates” for more detail.

2.Normal Mode Rejection Ratio, at 50 or 60 Hz ±0.1%. The NMRR for 400 Hz ±0.1% is 85 dB in S rate and 35 dB in M rate.

3.Common Mode Rejection Ratio at 50 or 60 Hz ±0.1%, with 1 kΩ in series with either lead. The CMRR is >140 dB at dc for all reading rates.

4.20 volts or 2 times full scale whichever is greater, not to exceed 1000V.

5.Reading rate-1/3 rdg / sec. in the 20 mV, 20Ω, 200 mA dc ranges

6.Reading rate-1.25 rdg / sec. in the 20 mV, 20Ω, 200 mA dc ranges

TRUE RMS AC VOLTAGE (OPTION 8842A-09)

 

 

Input Characteristics

 

 

 

 

 

 

 

 

 

 

 

RANGE

FULL SCALE 5ñ

 

RESOLUTION

 

 

DIGITS

 

 

 

 

 

 

5ñ DIGITS

4ñ DIGITS*

 

 

 

 

 

 

 

 

 

 

 

200 mV

199.999 mV

 

1 μV

10 μV

 

2V

1.99999V

 

10 μV

100 μV

 

 

 

 

100 μV

 

 

20V

19.9999V

 

1 mV

 

 

 

 

 

 

 

200V

199.999V

 

1 mV

10 mV

 

 

 

 

 

 

 

700V

700.00V

 

10 mV

100 mV

 

 

 

 

 

 

 

*4ñ digits at the fastest reading rate

 

 

 

 

 

 

 

 

INPUT

IMPENDANCE

1 MΩ

Shunted

By

<100 pF

1-4

Page 18
Image 18
Fluke 8842a instruction manual Rate, Filter, Peak NM Signal, True RMS AC Voltage Option 8842A-09, Input Impendance