8842A
Instruction Manual
3-17. G5 (Get IAB Status)
The G5 command loads the output buffer with the IAB status in the format shown in Figure
Example Explanation
1011 CR LF 1: Leading 1
0:FRONT inputs
1:Autorange off
1: OFFSET feature on
It is useful to know whether autorange is on or off because this information is not available from the G0 command. For example, the G0 command could indicate that the 8842A was in the 200 mV range, but it would not indicate whether the 8842A was in autorange or manual range.
3-18. G6 (Get YW Status)
The G6 command loads the output buffer with the YW status in the format shown in Figure
Example Explanation
1015 LF CR 1: Leading 1
0:Leading 0
1:Y1 (enable output suffix)
5:W5 (enable LF only)
3-19. G7 (Get Error Status)
The G7 command copies the error status register into the output buffer in the format shown in Figure
Example | Explanation |
1071 CR LF | 1: Leading 1 |
| 0: Leading 0 |
| 71: Syntax error in |
The G7 command gives the error status as it exists when the command is executed at its position in the input string. The G7 command does not clear the error status register. For more information about error messages, see paragraph