Fluke 8842a instruction manual Internal Fuse Replacement, Initial Troubleshooting Procedure

Models: 8842a

1 269
Download 269 pages 45.95 Kb
Page 173
Image 173

8842A

Instruction Manual

WARNING

TO AVOID ELECTRIC SHOCK, ENSURE THAT

THE GROUNDING SCREW IS FIRMLY

ATTACHED TO THE CASE BOTTOM.

6-34. INTERNAL FUSE REPLACEMENT

CAUTION

For fire protection, use exact fuse replacements only.

The 8842A has an internal 3A 600V slow-blow fuse (F301) in series with the 2A input terminal. To replace this fuse, remove the case according to the disassembly instructions. The fuse is held in fuse clips on the Main PCA. Do not use makeshift fuses or short- circuit the fuse holder.

6-35. EXTERNAL TRIGGER POLARITY SELECTION (Option -

05 Only)

The EXT TRIG input is factory-configured with negative polarity (trigger on falling- edge). This polarity is set by jumper E902 on the IEEE-488 Interface PCA. To select positive polarity (trigger on rising-edge), remove jumper E902 and add jumper E903.

6-36. TROUBLESHOOTING

The 8842A is designed to be easily maintained and repaired. Both the analog and digital circuits have built-in diagnostic self-tests and troubleshooting modes to facilitate troubleshooting and repair. The instrument’s circuits allow troubleshooting and repair with basic electronic troubleshooting equipment such as a multimeter and oscilloscope. The troubleshooting mode in the digital controller circuitry generates special test signals to allow troubleshooting and repair without a special test signal generator or complex logic analyzer. Using the information in this section, a technician should be able to troubleshoot and repair the 8842A very efficiently. There is also a troubleshooting package available which utilizes the Fluke 9010A System Troubleshooter. The 8842A- 9000 Troubleshooting Kit is described in detail in Section 8.

6-37. Initial Troubleshooting Procedure

WARNING

TO AVOID INJURY OR EQUIPMENT DAMAGE, USE EXACT

REPLACEMENT PARTS FOR ALL PROTECTION COMPONENTS.

When a problem occurs in the 8842A, first verify the problem is actually in the instrument. If the problem occurs when the instrument is in a system, check to see if the same problem exists when under local control. Watch the display as the instrument is turned on to see if any of the digital self-test error codes appear indicating a digital failure. If the malfunction does not involve the True RMS AC or IEEE-488 options, remove the option(s) from the instrument before proceeding.

6-36

Page 173
Image 173
Fluke 8842a instruction manual Internal Fuse Replacement, Initial Troubleshooting Procedure