Fluke 8842a instruction manual Reassembly Procedure, Front Panel Disassembly

Models: 8842a

1 269
Download 269 pages 45.95 Kb
Page 169
Image 169

8842A

Instruction Manual

6-32. Front Panel Disassembly

1.Holding the chassis vertically (with the front panel downward), remove the mounting screws from the four corners of the Display PCA (Figure 6-7A).

2.Holding the chassis vertically (now standing the instrument on the rear panel), pull the front panel off the chassis and set it aside (Figure 6-7B).

3.Remove the the display assembly (Display PCA, spacer matrix, and keypad) from the chassis as follows (see Figure 6-7C):

a.Release the two plastic tabs on the front of the chassis.

b.Let the bottom edge of the display assembly swing toward the rear of the instrument.

c.Pull the Display PCA toward the bottom of the chassis.

4.Separate the spacer matrix from the Display PCA by releasing the two pairs of plastic snap tabs on the back of the Display PCA (Figure 6-7D).

5.Remove the keypad from the spacer matrix.

CAUTION

The vacuum fluorescent display should not be removed from the Display PCA; these are supplied as one part.

6.Remove the display window from the front panel as follows:

a.Slide the window upward (away from the buttons) about 1/32 inch (Figure 6-8A).

b.Push the window directly outward from the front panel (Figure 6-8B).

6-33. REASSEMBLY PROCEDURE

To reassemble the instrument, proceed as follows:

1.Assemble the front panel assembly by reversing the front panel disassembly procedure (Figure 6-7). (It is easiest to lay the keypad on the Display PCA before installing the spacer matrix on the Display PCA.)

CAUTION

The four Display PCA mounting screws are self-tapping. To avoid damaging the threads, ensure the screws are threaded properly before tightening. Do not overtighten them.

2.Turn the chassis upside down.

3.Install the Main PCA through the bottom of the chassis as follows:

6-32

Page 169
Image 169
Fluke 8842a instruction manual Reassembly Procedure, Front Panel Disassembly