SPECIFICATIONS

Accuracy

 

NORMAL (s) READING RATE

±(% of Reading + Number of Counts).

 

 

For sinewave inputs 10,000 counts1.

 

 

 

FREQUENCY

20-45

45-200

200-20k

(200 mV range)

(2V-200V range)

(700V range)

20k-50k

50k-100k

24 HOURS2 23±1°C

1.2+ 100

0.3+ 100

0.06+ 100

0.05+ 80

0.06+ 100

0.15+ 120

0.4+ 300

90 DAY 23±5°C

1.2+ 100 0.35 + 100

0.08+ 100

0.07+ 80

0.08+ 100

0.19+ 150

0.5+ 300

1 YEAR 23±5°C

1.2+ 100

0.4+ 100

0.10+ 100

0.08+ 80

0.10+ 100

0.21+ 200

0.5+ 400

2 YEARS ±5°C

1.2+ 100

0.5+ 100

0.20+ 100

0.15+ 80

0.20+ 100

0.25+ 250

0.5+ 500

1.For sinewave inputs between 1,000 and 10,000 counts, add to number of counts 100 counts for frequencies 20 Hz to 20 kHz, 200 counts for 20 kHz, and 500 counts for 50 kHz to 100 kHz.

2.Relative to calibration standards.

MEDIUM AND FAST READING RATES

........In medium rate, add 50 counts to number of counts. In the

 

fast rate the specifications apply for sinewave inputs

 

1000 (4½ digit PRGH￿ FRXQWV and >100 Hz.

NONSINUSOIDAL INPUTS

For nonsinusoidal inputs 10,000 counts with frequency

 

components 100 kHz, add the following % of reading to

 

the accuracy specifications.

 

 

FUNDAMENTAL

FREQUENCY

45 Hz to 20 kHz 20 Hz

20 Hz to 45 Hz and 20

kHz to 50 kHz

Operating Characteristics

1.0 TO 1.5

0.05

0.2

CREST FACTOR

1.5 TO 2.0

0.15

0.7

2.0TO 3.0 0.3 1.5

MAXIMUM INPUT

700V rms, 1000V peak or 2 x 107 Volts-Hertz product

 

(whichever is less) for any range.

TEMPERATURE COEFFICIENT

............±(% of reading + Number of Counts) per °C, to 18°C and 28°C

 

to 50°C.

 

 

FOR INPUTS

20-20k

FREQUENCY IN HERTZ

20k-50k

50k-100k

10,000 counts

1,000 counts

0.019 + 9

0.019 + 12

0.021 + 9

0.021 + 15

0.027 + 10

0.027 + 21

COMMON MODE REJECTION

..............>60 dB at 50 or 60 Hz with 1 kΩ in either lead.