Coupling Factor

Function Group

Menu Map

Description

ALC

1

This softkey allows specification of the coupling factor of an external coupler/detector used to externally level the synthesizer output power. Negative coupling factor values are required for valid entry. See “Specifications” for the coupling factor range.

Programming Codes SCPI: POWer:ALC:CFACtor <num>[dB]~MAXimum~MINimum Analyzer: NONE

See Also AALC

“Externally Leveling the Synthesizer” in Chapter 1.

CW

Function Group

Menu Map

Description

FREQUENCY

2

This hardkey lets you select a synthesized continuous wave frequency. When you press Icw), the synthesizer stops sweeping (green SWEEP

LED off) and displays: --> CW: XXXXX MHz. Where XXXXX represents a frequency value. Use either the rotary knob, the step keys (with or without the left/right arrow keys), or the numerical entry keys with a terminator key to set the desired value. If a small change is desired, use the left and right arrow keys to underline the digit to be changed.

Programming Codes

See Also

SCPI:

FREQuency[:CW] <num>[freq suffix] or

MAXimum/MINimumlUPlDOWN

FREQuency:MODE CW

Analyzer: CW

C#/CF Coupled, (s), (j)

“CW Operation and Start/Stop Frequency Sweep” in Chapter 1. “Programming Typical Measurements” in Chapter 1.

C-14 Operating and Programming Reference

HP 8360

 

User’s Handbook