SCPI COMMAND SUMMARY
the sweep mode then the output level is controlled by the start, stop, center and span functions. If in the fixed power mode then the output is controlled by the POW C: LEVEL1 command. The *RST value is FIXed.
•POWer:OFFSet <num>[DB]lMAXimumlMINimumlUPlDOWN
•POWer:OFFSet? [MAXimumlMINimum]
Sets and queries the power offset. This function changes mapping of absolute power parameters on input to and output from the synthesizer. Changing this does not affect the output power of the synthesizer. Only the displayed parameters and query responses are changed. The equation implied by this is:
The entered or displayed power = Hardware Power + Offset After *RST, the value is 0.
•POWer:OFFSet:STATe ONlOFFlllO
•POWer:OFFSet:STATe?
Queries and turns the power offset off and on. After *RST, the setting is OFF.
•POWer:RANGe <num>[lvl suffix] lMAXimumlMINimumlUPlDOWN
•POWer:RANGe?
Sets and queries the setting of the power meter range. This is used
when the command POWer:ALC:SOURce issettoPMETer.
•POWer:SEARch ONlOFFlllOlONCE
•POWer:SEARch?
Sets and queries the power search switch. This has an interaction
with POWer:ALC:STATe as described below.
POWer:ALC:STATe | POWer:SEARch | Power Switch Action |
ON | ON | ALC is momentarily closed to |
|
| level at the requested power, |
|
| and then the modulator is set to |
|
| the same voltage in open loop |
|
| mode. This repeats automatically |
|
| any time that the power level or |
|
| frequency is changed. |
ON | OFF | Normal mode. |
ON | ONCE | Immediately performs a power |
|
| search. This leaves POWer:SEARch |
|
| in the ON position. |
OFF | not applicable | Modulator setting is |
|
| explicitly set by user. |
|
|
|
•POWer:SLOPe <num>[DB/freq suffix]lMAXimumlMINimumlUPlDOWN
•POWer:SLOPe? [MAXimumlMINimuml
Sets and queries the RF slope setting (dB per Hz).
| HP 8360 |
| User’s Handbook |