E

8360 Adrs

Function Group

Menu Map

Description

SYSTEM

8

This softkey lets you change the HP-IB address of the synthesizer. Enter the address desired using the numeric entry keys or the up/down arrow keys. The address value may be set between 0 and

30.The synthesizer stores the address value in non-volatile memory. The default address of the synthesizer is 19.

Programming Codes

See Also

SCPI: SYSTem:COMMunicate:GPIB:ADDRess

Analyzer: NONE

Connectors, HP-IBMenu

“Getting Started Programming,” in Chapter 1.

“Programming Typical Measurements,” in Chapter 1.

Enter Corr

Function Group

Menu Map

Description

POWER

5

This softkey lets you enter a power correction value for a frequency point in the flatness array. A frequency point must be entered before a correction value can be accepted, otherwise the following error message appears:

ERROR - Must first enter correction freq. The up/down arrow keys let you scroll through the frequency points available for power correction. If no correction value is entered for a frequency point, the synthesizer display indicates Undefined. The range of acceptable values is -40 to +40 dB. An asterisk next to the key label indicates that this feature is active.

HP 8360

Operating and Programming Reference E-l

User’s Handbook