SCPI COMMAND SUMMARY
a *OPT?
This returns a string identifying any device options.
0 *RCL <num>
The instrument state is recalled from the specified memory register.
The value range is from 0 to 8.
•*RST
The synthesizer is set to a predefined condition as follows:
AM:DEPTH value is 50%
AM INTernal FREquency value is 1 kHz
AM:MODE NORMal
AM:SOURce EXTernal
AM:STATe OFF
AM:TYPE LINear
CALibration:PEAKing:AUTO OFF
CALibration:POWer:ATTenationO DBM
CALibration:POWer:RANGel
CALibration:SPAN:AUTO OFF
CORRection:FLATness? returns a0
CORRection:ARRayclear
CORRection:FLATness:POINts? returns a0
CORRection:STATe OFF
DIAGnostics:ABUS:AVERage 1
DIAGnostics:TEST:ENABle ALL
DIAGnostics:TEST:LOG:SOURceFAIL
DIAGnostics:TEST:LOG[:STATel OFF
DIAGnostics:TEST:LOOP OFF
DISPlay[:STATel ON
FM:DEViationvalueis 1 MHz
FM:COUPlingAC
FM:FILTer:HPASs MAXimum
FM:INTernal:FREquency valueis 1 MHz
FM:SENSitivity MAXimum
FM:SOURce EXTernal
FM:STATe OFF
FREquency:CENTervalueis(MAX+MIN)/2
FREQuency:CWvalueis(MAX+MIN)/2
FREQuency:CW:AUTO OFF
FREQuency:MANualvalueis (MAX+MIN)/2
FREQuency:MODECW
FREQuency:MULTiplier 1
FREQuency:MULTiplier:STATe OFF
FREQuency:OFFSet 0
FREQuency:OFFSet:STATe OFF
FREQuency:STARtMINimum
FREquency:STEP calculated from span
FREQuency:STEP:AUTO ON
FREQuency:STOP MAXimum
HP 8360 | Operating and Programming Reference |
User’s Handbook |
|