Setup Synthesizer Parameters
6.On the synthesizer, press (PRESET).
7.FREQUENCY ISTART) @ LGHz), LSTOP) 0 @ LGHz).
8.(POWER LEVEL) (TJ m.
Access User Flatness Correction Menu
9.Press POWER [MENU). Select Fitness Menu.
10.Select Delete Menu Delete All . This step insures that the flatness array is empty.
11.Press (6%). Leave the delete menu and return to the previous soft key menu.
12Enter the frequency points at which the correction information will be taken. Choose either the
Auto Fill Start. For this example,select Auto Fill Start @IGHz).
13Select Auto Fill Stop a@=, Auto Fill Incr a[GHz).
Notice that a frequency list starting at 4 and ending at 10 GHz with an increment value of 1 GHz is created.
Enter Correction Data into Array
14.Select Mtr Meas Menu Measure Cars All. The power meter is now under synthesizer control and is performing the sequence of steps necessary to generate the correction information at each frequency point.
If an
Enable User Flatness Correction
15.When the operation is complete, (a message is displayed) the flatness correction array is ready to be applied to your setup. Disconnect the power meter/sensor and press [FLTNESS ON/OFF) (amber LED on). The power produced at the point where the power meter/sensor was disconnected is now calibrated at the frequencies and power level specified above.