Power Offset
Function Group P O W E R
Menu Map 5
Description This softkey changes the mapping of absolute power parameters on input to the synthesizer. It does not change the RF output produced by the synthesizer. The equation used to determine the displayed value is:
Entered or Displayed Power = Hardware Power (ALC) + Active Offset.
Programming Codes SCPI:
POWer:OFFset:STATe ON]OFF]llO
POWer:OFFset <num>[DB]]MAXimum(MINimum(UP(DOWN
Analyzer: NONE
See Also [POWER LEVEL) and POWER (FV1ENU).
Power Slope
Function Group POWER
Menu Map 5
Description This softkey lets you compensate for system, cable, and waveguide variations due to changes in frequency, by linearly increasing or decreasing power output as the frequency increases. RF slope values may range from
Programming Codes SCPI:
POWer:SLOPe:STATe ON(OFFIl]O
POWer:SLOPe <num>[DB](MAXimum]MINimum]UP]DOWN
Analyzer: SLl function on, SLO function off. Note that because SL functions in the fundamental units of dB/Hz, you program the SL code SLmdt, where m is 1 (on) or 0 (off); d is the numerical value
| HP 8360 |
User’s | Handbook |