9.Press (PRIOR). Leave the delete menu and return to the previous soft key menu.
10. Select Auto Fill Start @ m). Set the first frequency in correction table to 2 GHz.
11.Auto Fill Stop @ @J (GHz). Set the last frequency in correction table to 20 GHz.
12.Auto Fill Incr 0 @ @ INIHz). Set the frequency increment to every 100 MHz from 2 to 20 GHz.
Setup Power Meter
13.Zero and calibrate the power meter/sensor.
14.Connect the power sensor to test port.
15.Enter and store in the power meter, the power sensor’s cal factors for correction frequencies to be used.
Enter Correction Data into Array
16.Select Mtr Meas Menu Measure Corr All . The power meter
is now under synthesizer control and is performing the sequence of steps necessary to generate the correction information at each frequency point.
If an
Enable User Flatness Correction
17.When the operation is complete, (a message is displayed) the flatness correction array is ready to be applied to your setup.
18.Disconnect the power meter/sensor.
19.On the synthesizer, press [POWER LEVEL) 0 (dBm). Where
n = PO max - Ppath l o s s for maximum leveled power at the test port.
20.To save the synthesizer parameters including the correction table in an internal register, press ISAVE) (XJ.
(n = number 1 through 8).
Reactivate the HP 8757 System Interface
21.Set the analyzer to SYSINTF ON, the analyzer and synthesizer preset.
22.Press (RECALL) 0, Recall the synthesizer parameters from storage register 1