D

Dblr Amp Menu

Function Group POWER

Menu Map 5

Description This softkey accesses the doubler amp mode softkeys. These softkeys are applicable to instrument models with a doubler installed. The doubler has an integral amplifier whose operation is controlled by the instrument firmware. Its use depends on the frequency of operation and on the calibration constants set at the factory. The instrument defaults after preset to this automatic mode of operation which is the specified operation. Softkeys in this menu will allow you to turn the doubler amplifier always on or always off. These two modes are unspecified operation for instruments with a doubler installed. These softkeys have no effect on instruments without a doubler.

Doubler Amp Mode AUTO

Sets the doubler amp mode to

 

AUTO. This is the default after

 

preset and must be used for specified

 

performance.

Doubler Amp Mode On

Turns the doubler amplifier on

 

regardless of the frequency of

 

operation. Using this mode results in

 

unspecified performance.

Doubler &rip Mode Off

Programming Codes SCPI: NONE

Analyzer: NONE

Turns the doubler amplifier off regardless of the frequency of operation. Using this mode results in unspecified performance.

See Also Softkeys listed above.