DIODe Initiates a calibration of the external flatness. Depends on value of CALibrat ion: PMETer : RANGe.

PMETer Initiates a calibration of the power meter flatness. Depends on value of CALibrat ion : PMETer : RANGe.

MMHead Initiates a calibration of the source module flatness. Depends on value of CALibrat ion : PMETer : RANGe.

CALibration:PMETer:FLATness:NEXT? <num>[lvlsuffix]

The parameter is the measured power that is currently produced by the synthesizer. You must supply this parameter after measuring the power using an external power meter. The query response is issued after the synthesizer processes the supplied parameter and settles on the next point to be measured. The query response is:

>o 0

<O

The frequency [in Hz] that is currently produced. The calibration is complete.

An error has occurred and the calibration is aborted.

a CALibration:SPAN:AUTO ONlOFFlllO

CALibration:SPAN:AUTU?

Sets and queries the automatic sweep span calibration.

ON

A

calibration

is

done

whenever the sweep span is changed.

OFF

A

calibration

is

done

only when

 

CALibrat ion : SPAN [ : EXECuteI is sent.

After *RST, the setting is OFF.

CALibrat ion : SPAN [ : EXECut el Causes a sweep span calibration.

CALibrat ion : TRACk

Causes an automatic tracking calibration procedure.

CORRection:ARRay[il (<num> [DBI )1601*1601

CORRect ion : ARRay [il?

Sets and queries the entire 1601-point array of correction values that can be added to the internal flatness correction array. The

1601-points are added to the internal flatness array synchronized on the trigger output 1~s pulses. These TTL-level pulses are 1601 evenly spaced points across an analog sweep, or at each point in step or list mode. Entering this array causes the CORRection:SOURce Cil command to set to ARRay. There is one array for the foreground state (i=O) and one for the background state (i=l). If the [i] is not specified, the default value is i=O. After *RST, these arrays are cleared.

CORRect ion : FLATness (<num> [f req suffix] , <num> CDBI 32*801

CORRection:FLATness?

S-30 Operating and Programming Reference

HP 8360

User’s

Handbook