Fault Info 2

Programming Codes SCPI: NONE

Analyzer: NONE

See Also Fault Menu

Fault Info 3

Function Group SERVICE

Menu Map 6

Description This softkey displays the latched status of the following fault messages.

CALYO FAIL

Indicates that the YO adjusted at power-on or at

 

preset is unable to calibrate. Initiate a full self-test

 

to gather more information if this fault is indicated.

CALMAN

Indicates that the manual sweep DAC adjusted at

FAIL

power-on or at preset is unable to calibrate. Initiate

 

a full self-test to gather more information if this fault

 

is indicated.

TMR

Indicates a possible internal software error. Two

CNFLCT

routines are trying to use the same timer.

FAIL

 

SEARCH

Indicates that the ALC search leveling algorithm has

 

failed. This fault indication is possible only if the

 

search leveling mode is on.

Programming Codes SCPI: NONE

Analyzer: NONE

See Also Fault Menu

F-4 Operating and Programming Reference

HP 8360

 

User’s Handbook