Internal Pulse Generator Rate

Function Group IMOD)

Menu Map 4

Description This softkey (Option 002 only) lets you set a value for the internal pulse generator’s pulse rate. The range of acceptable values is from

2.5Hz to 3.33 MHz. (These values are obtained by taking the inverse of the period.) The factory preset default is 500 Hz. When this feature is active, its current value is displayed in the active entry area.

Since rate and period are inversely related, if both are given values, only the last one will be applied which will cause the first one to be recalculated. Use the one that is convenient for your application. For example, if you set the pulse rate, do not change the pulse period (the synthesizer automatically adjusts the period to match the rate.)

Programming Codes SCPI: PULM:INTernal:FREQuency <num>[freq suffix]JMAXimum~MINimum

Analyzer: NONE

See Also 0,MOD a 1 so see “Pulse” and “Modulation”.

Internal Pulse Generator Width

Function Group IlvloD]

Menu Map 4

Description This softkey (Option 002 only) lets you set a value for the internal pulse generator’s pulse width. The pulse is adjustable from 25 ns to 400 ms with 25 ns resolution. The factory preset default is 1 ms pulse width. If you set a value for the pulse width that is greater than the pulse period, the pulse period is recalculated to a value equal to the pulse width plus 25 ns. When this feature is active, its current value is displayed in the active entry area.

I-10 Operating and Programming Reference

HP 8360

 

User’s Handbook

Page 235