Programming Codes SCPI: FREQuency:STEP[:INCR] <num>[freq suffix] or MAXimum I MINimum

Analyzer: SF or SHCF <num> [HzIKz~MzIGz]

See Also Manual Sweep, Sweep Mode Step, Up/Dn Size Swept

Up/Dn Size Swept

Function Group

Menu Map

Description

FREQUENCY

2

This softkey sets the frequency step size in the swept frequency step mode. The step size may be set from 1 Hz to 10 GHz. The factory preset step size is 100 MHz. Step size values are entered using the entry area.

If an underline cursor appears under a digit in the entry display, then the value will be modified by the up/down arrow keys or the rotary knob. The increment/decrement size in this case is the underlined digit by the power of 10.

If the up/down function is on (asterisk next to key label) and the cursor is not under one of the active entry area digits, then frequency value is changed by the up/down size using either the up/down arrow keys or the rotary knob.

Programming Codes SCPI: FREQuency:STEP[:INCR] <num>[freq suffix] or MAXimumlMINimum

Analyzer: SF or SHCF <num> [HzlKzlMzlGz]

See Also Up/Dn Size CW

HP 8360

Operating and Programming Reference U-3

User’s Handbook

 

Page 383
Image 383