•ABORt
Causes the sweep in progress to abort and reset. If INIT: CONT is ON it immediately restarts the sweep. The pending operation flag (driving *OPC, *WAI, and *OPC?) undergoes a transition once the sweep is reset.
•AM[:DEPTh] Qnun>[PCTlIMAXimumlMINimuml~num>DB
•AM[:DEPThl?[MAXimumlMINimum]
Sets and queries the percentage of AM when the SOURce is INTernal. If <num> is received with units of dB, the value is converted to percent by the equation:
[%] = 100 * (1 -
Valid ranges of dB are 0 to 40 dB. After “RST, the value is 50%.
•AM:INTernal:FREQuency Qnun>[Cfreqsuffix>]lMAXimumlMINimum
•AM:INTernal:FREQuency?[MAXimumIMINimum]
Sets and queries the frequency (in Hz) of the internal AM source.
After *RST, the value is 1 kHz.
•AM:INTernal:FUNCtion SINusoidlSqUarelTRIanglelRAMPlNOISe
•AM:INTernal:FUNCtion?
Sets and queries the waveform of the internal AM source. After “RST, the value is SINusoid.
•AM:SOURce INTernallEXTernal
•AM:SOURce?
Sets and queries the source of the AM modulating signal. After *RST, the value is EXTernal.
•AM:MODE DEEPlNORMal a AM:MODE?
Controls the AM depth limits of the synthesizer. The NORMal position is selected at *RST.
•AM:STATe ONlOFFlllO
•AM:STATe?
Sets and queries the status of the AM modulation. After *RST, the
setting is OFF.
•AM:TYPE LINearlEXPonential a AM:TYPE?
Sets and queries the type of AM modulation. After *RST, the setting is LINear.
•CALibration:AM:AUTO ONlOFFlllO
a CALibration:AM:AUTO?
Sets and queries the automatic modulator calibration switch.
| HP 8360 |
User’s | Handbook |