Preset ModeFactory

Function Group

Menu Map

Description

Preset'lfade Factory

SYSTEM

8

This softkey sets the standard starting configuration of the synthesizer when the (PRESET) key is pressed, as set by the manufacturer. An asterisk next to the key label indicates that this feature is active. The following is a description of the configuration.

nStart sweep at the minimum specified frequency. H Stop sweep at the maximum specified frequency.

nPower level set at 0 dBm.

nSweep time set to auto.

nCONT sweep.

w Sweep mode ramp.

nALC leveling point internal.

nALC leveling mode normal.

nMarkers set to activate at the center frequency of the sweep.

nAll function values stored in memory registers 1 through 9 remain in their previous states.

nThe checksum of the calibration data is calculated, and if an error is detected, the calibration data in protected memory is used. If the checksum of the protected data is not correct, then default values are used an error message (EEROM FAILED, LOST CAL) is displayed.

Programming Codes SCPI: SYSTem:PRESet:TYPE FACTory Analyzer: IP, which is the same as cm).

See Also 07P R E S E T Preset Mode User

“Changing the Preset Parameters,” in Chapter 1.

HP 8360

Operating and Programming Reference P-9

User’s Handbook