How to View or Change an HP-IB address from the Front Panel
To set an
1.Press SYSTEM (e).
2.Select
3.The synthesizer displays the three address softkeys: 8360 A&s, Meter Adrs , and Printer Adrs .
4.Select the desired softkey.
5.The synthesizer displays the address selected for that instrument.
6.If you want to change the address, use the keypad to enter the desired address (0 to 30), then press @%?j.
If the synthesizer displays Rear panel
How to Prevent a Front Panel Change to an HP-IB Address
To disable the address softkeys, set the instrument address on the rear panel
How to Set the
If your synthesizer does not have a front panel, set the address on the rear panel
All of the externally mounted connectors on the instrument are discussed in the “Operating and Programming Reference” chapter, under “Connectors.” If you are interested in the HP part number for a connector, see “Replaceable Parts” in this manual set.
To keep the internal timebase frequency reference oven at operating temperature, the synthesizer must be connected to ac line power. The synthesizer requires approximately 30 minutes to warm up from a cold start before the OVEN display message goes off. With
a stable outside temperature, internal temperature equilibrium is reached after approximately two hours. For additional information on warmup times, see “Specifications,” in this volume.