HP 8360
User’s Handbook
SCPI COMMAND SUMMARY
FREQ:MODE Affect on Slope
CW or LIST Rotates around 0 Hz.
SWEep or Rotates around the start frequency.
STEP
The *RST value is 0.
•POWer:SLOPe:STATe ONOFFlO
•POWer:SLOPe:STATe?
Sets and queries the power slope state. *RST value is 0.
•POWer:SPAN <num>[DB] lMAXimumlMINimumlUPlDOWN
•POWer : SPAN? [MAXimum
I MINimum]
The coupling equations for power sweep are exactly analogous to those for frequency sweep. Power sweep is allowed to be negative, unlike frequency sweeps. *RST value is 0.
•
•POWer:STARt? [MAXimumlMINimuml
Default units and units for query response are determined by the command UNIT : POWer. The coupling equations for power sweep are exactly analogous to those for frequency sweep. Power sweep is allowed to be negative, unlike frequency sweeps.
*RST value is 0 dBm.
•POWer:STATe ONOFFlO
•POWer:STATe?
Sets and queries the output power on/off state. *RST value is OFF.
•POWer:STEP:AUTO ONOFFlO
•POWer:STEP:AUTO?
Sets and queries the function switch that controls how the power step size (Power: STEP: INCRement) is determined. if in the automatic state, then the step size is 1 dB. The *RST setting is ON.
•POWer : STEP [ : INCRementI <nu.m> [DB]
I MAXimum
1 MINimum
•POWer : STEP [ : INCRementI ? [MAXimumIMINimumI
Sets and queries the power step size to be used for any node in the power subsystem that allows UP and DOWN as parameters. Setting this value explicitly causes POWer : STEP : AUTO OFF.
The *RST setting is 10 dB.
•POWer:STOP Qn.un>[lvlsuffix] lMAXimumlMINimumlUPlDOWN
•POWer : STOP? [MAXimumIMINimumI
Sets and queries the ending power for a power sweep. Default units and units for query response are determined by the command UNIT:POWer. The coupling equations for power sweep are exactly