Step Control Master

SCRLAR

NRSTER

S L R V E

NETUORK RNRLYZER

S Y N T H E S I Z E R

S Y N T H E S I Z E R

Figure S-l. Connections Required for a Two-Tone Scalar Network Analyzer Measurement System

1.Designate one synthesizer as the master, the other as the slave.

2.Make the connections.

3.To avoid synchronization problems, always set up the slave (frequency and power) before setting up the master.

4.Set up the master (frequency, power, and sweep time).

5.Set the sweep time on the slave.

6.Configure the synthesizers for step sweep, or ramp sweep.

7.Select the appropriate triggering scheme.

8.Activate the slave mode on the slave synthesizer.

9.Activate the master mode on the master synthesizer.

By connecting the master’s 10 MHz reference standard to the slave’s 10 MHz reference input, the master synthesizer’s timebase supplies the frequency reference for both synthesizers.

In step sweep measurements, if the master synthesizer is not connected to an external controller, it must automatically trigger the steps. Set Step Sq Pt Trig Auto on the master. When a the scalar network analyzer is the step sweep controller, set Step Sup P% Trig Bus on the master synthesizer so that the analyzer can trigger the steps.

Programming Codes SCPI:

SWEep:CONTrol:STATe ONlOFFlllO

SWEep:CONTrol:TYPE MASTer

Analyzer: NONE

See Also Step Control Slave, Step Swp Menu

HP 8360

Operating and Programming Reference S-65

User’s Handbook

 

Page 363
Image 363
HP 8360 manual See Also Step Control Slave, Step Swp Menu