Tr o u b l e s h o o t i n g

Characterization Failure

Repeat the characterization making sure that the:

Room is dimly lit or a dark cloth is placed over the monitor

Sensor is securely attached to the monitor if the sensor has suction cups

Enable cable and the sensor connections are properly made

Dark Gain Correction of the sensor is performed correctly

Monitor is warmed up properly

The monitor brightness may be set too low. Use the factory default setting. If that fails, increase the brightness.

Verification Failure

Repeat the calibration and characterization making sure that the:

Room is dimly lit or a dark cloth is placed over the monitor

Sensor is attached to the monitor securely if the sensor has suction cups

Enable cable and sensor connections are properly made

Dark Gain Correction of the sensor is performed correctly

Monitor is warmed up properly

Refer to the Verification logs. See “Verification Logs” on page 73.