Tektronix 494A Characteristic Performance Uirement, Normal Accuracyand Ldentical Resolution

Models: 494AP 494A

1 336
Download 336 pages 53.32 Kb
Page 30
Image 30

Specilication - 494A/494Ap Service, Vot. 1

 

 

 

Table 2-l

(Continued)

 

 

FREQUENCY

 

 

 

RELATEDCHARACTERTSTICS

Characteristic

Performance

uirement

SupplementalInformation

Line-relatedSidebands

 

 

cally (-55 dBc (47 Hz - 44A Hz)

EffectiveVideo Bandwidth

 

 

 

kHz

Pulse StretcherFall Time

Marker(s)

3 kHz

300 Hz

30 Hz

3 H z

0.3Hz

Typically 30 psldivision of pulse tude

When activated, the marker is a bright dot positioned by the

CENTER/MARKER FREQUENCY control or the DATA ENTRY pushbut- tons.

Normal

 

 

Accuracyand

ldentical

 

Resolution

For the active trace

 

to centerfrequency

Delta Marker

 

 

Accuracy

t1o/o of the tolal span

For the active trace.

Resolution

FrequencySpan/Div OverallRange

MinimumSpan/Div

5/" at the measurernent on multi- band and stored displays

Displays delta time in Zero Span mooe

AMKR activates a second marker at the position of the single marker on the trace. Parenthesisappear on the marker displayline indicatingthat the delta mode is active. The display shows the difference in frequency and amplitude. 1-MKR-2 selects which marker is tuned.

At least10/ ot Span/Div

10Hzldivto 10 GHz/div(in a 1-2-5 sequencewith SPAN/DIVcontrol)or

10Hzldiv to 15 GHz/div(from the DATA ENTRY pushbuttons)to two significantdigits.

10 Hz in all bands