MANUAL CHANGE INFORMATION

Date: 5/u9o

ChangeReference:C2lSgO

 

Product:494N4g4Ap ServiceVol. 1

 

ManualPart No. 070-5s60-OO

product Group: 2E

 

 

In section4 - PERF)RMANCE)HECK pRocEDURE change stepr r as foltows:

Replace header secfian with the foltowing:

11. Check ResolutionBandwidthand Shape Factor (bandwidthis within2oo/oot the selectedLandwidthfor allbutthe'l0Hzfilter;.10Hzto 1 MHz in decadesteps, and 3 MHz; shape factor is 7.5:.t or less _ 60 dB bandwidthfor the .t0Hz filreris s150 Hz)

Change step i. to read as follows:

.i. Repeatthe processto checkthe resolutionbandwidth and shapelactorfor the 10 kHz, 1 kHz, and 100 Hz fihers.

j.To checklhe60dBbandwidthofthe 10 Hzfitter,setthe SpanrDivro 50 Hz, and the REFEHENCEt_EVfL lo _20 dBm. Push AUTO RES, VTDEOFTLTERWlOe, and ser PEAK AVERAGEto average.

k. Checkthatthe 60 dB bandwidthis <150 Hz.

Page 325
Image 325
Tektronix 494AP service manual Date 5/u9o, Perfrmanceheck pRocEDURE change stepr r as foltows, Product494N4g4Ap ServiceVol