PerforrnanceCheck procedure- 4g4ful4g4ApServiceVol. 1

33. Check Option 4t

Frequency Span/Div Accuracv

!1t73 o.t5 MHz/Divover the cdnter6 divisionsof the display)

., Sp:laccuracy is c_heckedat centerfrequencyset- tingsof 6 GHzand11 GHz.

a. 'Connect

the test equipmentas shownin Figure

4-26. set the sp€ctrumAnalyzer follows: "ontroir ",

FREQRANGE5.4-18GHz

CENTERFREQUENCY 6 GHz

sPAN/DtV5 MHz

RESOLUTIONBANDWIDTH 100kHz

REFLEVELAs Needed

TrMElDtvAUTO

VERTICALDISPLAY 10dB/Dtv

b.Modulatethe Comb Generatorsignal with .2ps timemarkers.

c.peak the responsewith the MANUALpEAKcon- trol andset REFLEVELfor the best markerdefinition.

d.use the HorizontalposlTloN control on the Spectrum Anafyzer to position a marker to center screenthen checkthe accuracyoverthe centersix divi- sionsof the disptay.

e.Check that.the^tim€marks align with the maior graticulelineswithinS0kHz.

f.RESETth€ CENTERFREQUENCYtO11 GHZ.

S.Check SPAN/DV accuracy. Error must not exceed*50 kHz/Div.

SPECTRUTANLAYZERUIiIDERTEST

COMB

GENERATOR

MOOULE

Figure4-26. Testequipmenisetupfor checkingoption 4i span/Divaccuracy.