Status reporting | IEEE 488.2 compatible Status Structures |
6Status reporting
IEEE 488.2 compatible Status Structures
Power
Modules
Status
Structure
The Power Modules Status Structure is read with one Query: CSTS? <ch>. The returned status contains the Events Register, Warnings Register, Output Register, Faults Register, Status Register and Error Code.
The Events Register
PON
0
CMD
OUT
ERR
FLT
WRN
OPC
PON | set when a Power On event occurs. |
CMD | set when the Power Module detected an error while communicating with |
| the XMP 2600 Main Controller.. |
OUT | the Output Register had a bit changed. |
ERR | set when an Error event occurred. |
FLT | the Faults Register had a bit changed. |
WRN | the Warnings Register had a bit changed. |
OPC | Operation Complete - the Power Module finished processing the XMP |
| 2600 Main Controller Command. For commands that effect the output of |
| the Power Module, the OPC flag is set when the output settled to within |
| 2% of full scale from the programmed value. |
The Events Register shows events that occurred in the Power Module. Events are changes in the Power Module’s status. The register is cleared when read.
The Warnings Register
0
CONF
FOLD
V / C
LC
LV
HC
HV
CONF | the Confidence Test (BIT) found errors. |
FOLD | Foldback is active. |
V / C | Voltage (0) or Current (1) Mode. |
LC | Low Current Window Warning. |
LV | Low Voltage Window Warning. |
HC | High Current Window Warning. |
HV | High Voltage Window Warning. |
The Warnings Register shows the current state of the Power Module.
XMP 2600 Programming Manual rev. 1.1 | 69 |