5. DESCRIPTION OF EACH PART

Please refer to Chapter 14 specification attached at the end of the manual.

6.PREPARATION FOR OPERATION

6.1Injection of clean water

Inject clean water through the filter or air aspiration port until it overflows from the gas conditioner.

6.2Supply of external power

Supply external power to the analyzer. And check the connections with output signal wire, etc. The exter-

nal power supply should be.........

100V± 10% AC, 50/60 Hz ± 1%,

with power consumption of 0.9kVA max. (excluding heating pipe). Check with reference to the separately issued approval diagrams.

6.3Check of sample piping, exhaust piping, etc.

Check the connection of the sample piping and if necessary, provide exhaust piping + drain piping (Pipes with inner diameter of ø8 or more for both). When sample will be dispersed upward, then branch and water- seal the exhaust pipe for the purpose of drain separation. Be sure to disperse the exhaust into the atmo- sphere.

Make checks with reference to the piping instructions in Fig. 4-1 and the separately issued gas analyzer diagram.

– 5 –

Image 13
Contents Infrared GAS Analyzer Request IntroductionProhibition Iii Prohibition Others Contents Vii Outline SpecificationsComposition PreparedSelection of installation site InstallationSelection of gas sampling location Connection of cables Installation workPiping procedure Supply of external power Injection of clean waterDescription of Each Part Preparation for OperationAirtightness test Setting of temperature inside lockerWarm-up time SO2 INDICATION/OPERATION PanelThis panel consists of the following functions Function and indication of indication/operation panelGeneral operation of indication/operation panel Advice on OperationKinds of standard gas CalibrationPreparation of standard gas Pressure regulation for standard gas cylinder Func CAL SET ENT Calibration concentration setting except zirconia O2 sensorCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operation Operating procedureRoutine maintenance and inspection MaintenanceConditioner Other maintenance and inspection Periodic maintenance and inspectionCentration is measured in the setting range of 0 to 200ppm or more at 0.5L/minGas extractor type ZBAK1 Maintenance of sampling deviceMembrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Others, major component figures Maintenance of zirconia O2 sensorStandard accessories Standard Accessories and SparesCode for preparation of 1-year spares and supplies Spares for 1 year optionalOutline Processing of Heating Pipe EndsEnd processing method Grid 300mm 40mm Grid area 150 mm Clamp support Check after pipingSpecifications of heating pipe Specifications of tube traceTroubleshooting of instrument TroubleshootingError codes and remedies Power supply Ambient temperaturePower consumption Mass weightMonth/NO2, 100ppm Standard requirements For sample gasNumber of output points and kinds of outputs Moving average output Block diagramOrdering information Automatic calibrationCalibration tolerance Automatic calibration output hold Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For indoor installation type For outdoor installation typeOnly specified outputs are wired Connection DiagramHead Office Instrumentation Div International Sales Dept