Fuji Labs ZSP instruction manual Func Hold ENT

Page 32

8.7.5Setting of output hold at calibration

This is used when calibrating for holding the output sig- nal at the value just be before entering calibration. (The indication is not held.)

The hold function is settable for each component.

Press the FUNC key in the measuring status and

’ will be indicated.

The HOLD LED of the function indicating lamps will flicker.

Press the key and the hold function will be settable. The sub indication will flicker.

Select hold ‘’ or ‘ ’ by pressing the key.

Press the ENT key after setting the hold function. Set- ting operation is completed.

Setting is valid for all the components.

At hold ON, the HOLD of the function indicating lamps will light steadily while the hold function is activated. At hold OFF, these lamps will go off.

In case of automatic calibration, output is held while the automatic calibration lamp is lit.

In case of manual calibration, output is held for about half an hour after completion of calibration.

For canceling the output hold forcibly, turn off the hold setting once.

FUNC

NO

HOLD

NO

NO

ENT

NO

Note 1: When measuring 4 components, or NOx, SO2, CO and O2 (with 2 infrared gas analyzers), output is held by using the external hold function at the time of automatic calibration.

Note 2: In the setting mode, output is held indifferently to the ON/OFF status of hold setting.

Note 3: In case the analyzer is equipped with a maintenance switch, output is held while the switch is turned on.

– 24 –

Image 32
Contents Infrared GAS Analyzer Introduction RequestProhibition Iii Prohibition Others Contents Vii Specifications OutlineComposition PreparedSelection of gas sampling location InstallationSelection of installation site Installation work Connection of cablesPiping procedure Injection of clean water Supply of external powerDescription of Each Part Preparation for OperationWarm-up time Setting of temperature inside lockerAirtightness test INDICATION/OPERATION Panel SO2Function and indication of indication/operation panel This panel consists of the following functionsAdvice on Operation General operation of indication/operation panelPreparation of standard gas CalibrationKinds of standard gas Pressure regulation for standard gas cylinder Calibration concentration setting except zirconia O2 sensor Func CAL SET ENTCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operating procedure OperationMaintenance Routine maintenance and inspectionConditioner Periodic maintenance and inspection Other maintenance and inspectionCentration is measured in the setting range of 0 to 200ppm or more at 0.5L/minMaintenance of sampling device Gas extractor type ZBAK1Membrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Maintenance of zirconia O2 sensor Others, major component figuresStandard Accessories and Spares Standard accessoriesSpares for 1 year optional Code for preparation of 1-year spares and suppliesProcessing of Heating Pipe Ends OutlineEnd processing method Grid 300mm 40mm Grid area 150 mm Check after piping Clamp supportSpecifications of tube trace Specifications of heating pipeTroubleshooting Troubleshooting of instrumentError codes and remedies Ambient temperature Power supplyPower consumption Mass weightNumber of output points and kinds of outputs Standard requirements For sample gasMonth/NO2, 100ppm Block diagram Moving average outputAutomatic calibration Ordering informationCalibration tolerance Automatic calibration output hold Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For outdoor installation type For indoor installation typeConnection Diagram Only specified outputs are wiredHead Office Instrumentation Div International Sales Dept