Fuji Labs ZSP instruction manual Specifications of tube trace, Specifications of heating pipe

Page 50

12.5 Specifications of tube trace

(1) Structure of tube trace

41

Finish outer diameter 33 Outer diameter of tube

10 mm

Heater cable

PFA teflon

3 2

5

qHeater cable (power limiting cable)

wProcess tube

eAluminum tape

rFiber glass heat insulator

tPVC jacket

 

 

 

Fig. 12-9

(2) Specifications of tube trace

 

 

 

 

 

 

q

Heater cable

HPT 20-1 BN 35.4w/m at 100ºC (100V AC)

 

 

 

 

 

w

Process tube

10/8 mm øPFA Teflon tube

 

 

 

 

 

e

Aluminum tape

50% overlap

 

 

 

 

 

r

Fiber glass

Thermal conductivity 0.2kJ/m·h·ºC

 

 

 

 

 

t

PVC jacket

UL 105ºC, black PVC

12.6Specifications of heating pipe

Maker: Thermon Far East

Dealer: Fujikin, Japan

– 42 –

Image 50
Contents Infrared GAS Analyzer Introduction RequestProhibition Iii Prohibition Others Contents Vii Composition SpecificationsOutline PreparedSelection of gas sampling location InstallationSelection of installation site Installation work Connection of cablesPiping procedure Description of Each Part Injection of clean waterSupply of external power Preparation for OperationWarm-up time Setting of temperature inside lockerAirtightness test INDICATION/OPERATION Panel SO2Function and indication of indication/operation panel This panel consists of the following functionsAdvice on Operation General operation of indication/operation panelPreparation of standard gas CalibrationKinds of standard gas Pressure regulation for standard gas cylinder Calibration concentration setting except zirconia O2 sensor Func CAL SET ENTCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operating procedure OperationMaintenance Routine maintenance and inspectionConditioner Centration is measured in the setting range of 0 to Periodic maintenance and inspectionOther maintenance and inspection 200ppm or more at 0.5L/minMaintenance of sampling device Gas extractor type ZBAK1Membrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Maintenance of zirconia O2 sensor Others, major component figuresStandard Accessories and Spares Standard accessoriesSpares for 1 year optional Code for preparation of 1-year spares and suppliesProcessing of Heating Pipe Ends OutlineEnd processing method Grid 300mm 40mm Grid area 150 mm Check after piping Clamp supportSpecifications of tube trace Specifications of heating pipeTroubleshooting Troubleshooting of instrumentError codes and remedies Power consumption Ambient temperaturePower supply Mass weightNumber of output points and kinds of outputs Standard requirements For sample gasMonth/NO2, 100ppm Block diagram Moving average outputCalibration tolerance Automatic calibration output hold Automatic calibrationOrdering information Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For outdoor installation type For indoor installation typeConnection Diagram Only specified outputs are wiredHead Office Instrumentation Div International Sales Dept