Fuji Labs ZSP instruction manual Auto CAL

Page 29

8.7.2Auto calibration start time setting

When FUNC key is pressed at the current time indica-

tion, the auto calibration start time and day will appear on the main and sub indications.

The AUTO CALLED of the function indicating lamps will flicker. In the case of start time, the decimal point on the main indication will light steadily.

Press the key and the auto calibration start time is settable, so set a start time on the main indication.

The numeric value is incremented by pressing the key.

Press the key and the digit can be selected.

After setting the calibration start time, press the key and the calibration start day can be set on the sub indica- tion.

The bar indication will flicker.

Press the key and the bar will shift.

The auto calibration start time is settable up to one week ahead.

When the auto calibration start time and day have been set, press the ENT key. The data are saved.

When the auto calibration at the set time is fin- ished, the next auto calibration start time will be set automatically.

To confirm the next auto calibration start time, then carry out this operation and check the indication.

SU TU TH SA

NO

MO WE FR

Lights

FUNC

SU TU TH SA

NO

MO WE FR

AUTO CAL

()

SU TU TH SA

NO

MO WE FR

( )

SU TU TH SA

NO

MO WE FR

ENT

SU TU TH SA

NO

MO WE FR

Note: When the auto calibration start time has been set, then set the auto calibration cycle.

– 21 –

Image 29
Contents Infrared GAS Analyzer Request IntroductionProhibition Iii Prohibition Others Contents Vii Outline SpecificationsComposition PreparedSelection of gas sampling location InstallationSelection of installation site Connection of cables Installation workPiping procedure Supply of external power Injection of clean waterDescription of Each Part Preparation for OperationWarm-up time Setting of temperature inside lockerAirtightness test SO2 INDICATION/OPERATION PanelThis panel consists of the following functions Function and indication of indication/operation panelGeneral operation of indication/operation panel Advice on OperationPreparation of standard gas CalibrationKinds of standard gas Pressure regulation for standard gas cylinder Func CAL SET ENT Calibration concentration setting except zirconia O2 sensorCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operation Operating procedureRoutine maintenance and inspection MaintenanceConditioner Other maintenance and inspection Periodic maintenance and inspectionCentration is measured in the setting range of 0 to 200ppm or more at 0.5L/minGas extractor type ZBAK1 Maintenance of sampling deviceMembrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Others, major component figures Maintenance of zirconia O2 sensorStandard accessories Standard Accessories and SparesCode for preparation of 1-year spares and supplies Spares for 1 year optionalOutline Processing of Heating Pipe EndsEnd processing method Grid 300mm 40mm Grid area 150 mm Clamp support Check after pipingSpecifications of heating pipe Specifications of tube traceTroubleshooting of instrument TroubleshootingError codes and remedies Power supply Ambient temperaturePower consumption Mass weightNumber of output points and kinds of outputs Standard requirements For sample gasMonth/NO2, 100ppm Moving average output Block diagramOrdering information Automatic calibrationCalibration tolerance Automatic calibration output hold Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For indoor installation type For outdoor installation typeOnly specified outputs are wired Connection DiagramHead Office Instrumentation Div International Sales Dept