7.INDICATION/OPERATION PANEL
7.1Name and description of each component of indication/operation panel
2 Main indication
1 Component indication
| 
 | 
 | 
 | 
 | 
 | 
 | Vol% | SU TU TH SA | 
 | |
| 
 | 
 | 
 | NOX | 
 | 
 | 
 | 
 | RANCE | |
| 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | ||
| 
 | 
 | 
 | 
 | 
 | ppm | 
 | 
 | 
 | |
| 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |
| 
 | 
 | 
 | 
 | 
 | 
 | 
 | MO WE FR | 
 | |
| 
 | 
 | 
 | SO2 | 
 | 
 | Vol% | 
 | 
 | RANCE | 
| 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | ||
| 
 | 
 | 
 | 
 | 
 | ppm | 
 | 
 | 
 | |
| 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |
| 
 | 
 | 
 | O2 | 
 | 
 | Vol% | 
 | 
 | RANCE | 
| 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | ||
| 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |
| 7 | Function indicator lamp | HEAS | CAL SET | SET | HOLD | 
 | RANCE | AUTO SAL | |
| 
 | 
 | 
 | FUNC | COMP | 
 | ENT | ZERO | SPAN | CAL | 
| 
 | 8 | Function key | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
| 
 | 9 | Component selector key | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
| 
 | 10 | Digit shift key | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
| 
 | 11 | Numeric input key | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
3Unit indication lamp
4Day indication
5Sub indication
6Range changeover switch
15Calibration start key
14Span calibration key
13Zero calibration key
12ENT key
| 
 | 
 | (Three component of NOx, SO2 and O2 are indication in this figure.) | |||||||||||||
| 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||
| Part Name | 
 | 
 | 
 | 
 | Description | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||
| 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||
| q Component indication | Indicates kind of gas measured. | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||
| w Main indication | Indicates measured concentration. Also indicates various  | ||||||||||||||
| 
 | calibration function (option), etc. | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||
| e Unit indication lamp | Indicates units of measured gas concentration. | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||||
| r Day indication | Indicates current day or day of starting by means of bar in auto calibration (option) setting mode. | ||||||||||||||
| 
 | Indication | SU | MO | TU | WE | TH | FR | SA | |||||||
| 
 | Day Sun. | Mon. | Tue. | Wed. | Thu. | Fri. | Sat. | 
 | 
 | 
 | 
 | 
 | 
 | ||
| t Sub indication | Indicates measuring range, error code, various set points, etc. | 
 | 
 | 
 | 
 | 
 | 
 | ||||||||
| y Range changeover key | Used when changing the range. High range is set when pressing | and low range is set when | |||||||||||||
| 
 | pressing . | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
| u Function indicator lamp | Relevant lamp lights up when following functions are set. | 
 | 
 | 
 | 
 | 
 | 
 | ||||||||
| MEAS: | Lights up in measuring status. | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||
| CAL SET: | Flashes in calibration concentration setting mode. | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||||||
| ALM SET: | Flashes in alarm setting mode. | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||
| HOLD: | Flashes in hold setting mode or lights steadily while hold function is activated. | ||||||||||||||
| RMT RANGE: | Flashes in remote range setting mode or light steadily while remote range function is activated. | ||||||||||||||
| AUTO CAL: | Flashes in auto calibration setting mode or lights steadily while auto calibration function is | ||||||||||||||
| 
 | activated. | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
| i Function key | Setting mode is changed at each press of this key (Refer to Section 5). | ||||||||||||||
| o Component selector key | Set component is changed for each setting mode. | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||||||
| !0Digit shift key | Shift is made from highest toward lowest digit at each press of this key. | ||||||||||||||
| !1Numeric input key | Selected digit is incremented at each press of this key. | 
 | 
 | 
 | 
 | 
 | 
 | 
 | |||||||
| !2ENT key | By pressing this key after setting, the set contents are memorized and becomes valid. | ||||||||||||||
| !3Zero calibration key | Used for zero point calibration. (Lamp flashes in zero calibration mode.) | ||||||||||||||
| !4Span calibration key | Used for span calibration (Lamp flashes in span calibration mode.) | ||||||||||||||
| !5Calibration start key | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | ||||
| Start key for manual calibration. Zero is calibrated by pressing | 
 | 
 | ZERO | and | CAL | keys. | |||||||||
| 
 | Span is calibrated by pressing | SPAN | and | CAL | keys. (CAL lamp lights steadily during calibra- | ||||||||||
| 
 | tion). | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
| 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
 | 
– 7 –