Fuji Labs ZSP Calibration concentration setting except zirconia O2 sensor, Func CAL SET ENT

Page 20

8.3Calibration concentration setting (except zirconia O2 sensor)

Set a span value for the calibration concentration setting. (Zero point calibration concentration is fixed at zero.)

Set the calibration gas concentration (span value). When FUNC key is pressed in the measuring sta- tus, the previously set span value will appear on the main indication.

The CAL SET LED of the function indicating lamps will flicker.

By pressing the key, the highest digit of the main indication for the 1st component will flicker, and the span value can not be set.

Select a range with the keys.

After selecting the range, set a span value in this status.

The numeric value will be incremented by press-

ing key.

The digit to be set can be selected by pressing key.

For selecting the 2nd and subsequent compo-

nents and the range, press the key and while the main indication for the 1st compo- nent is flickering, press the COMP key and then the main indication for the component to

be set will flicker. Now press the keys to select the range to be set.

When the span value has been set, press the ENT key. Setting operation is completed.

NO

FUNC

NO

CAL SET

NO

NO

ENT

NO

– 12 –

Image 20
Contents Infrared GAS Analyzer Introduction RequestProhibition Iii Prohibition Others Contents Vii Specifications OutlineComposition PreparedSelection of gas sampling location InstallationSelection of installation site Installation work Connection of cablesPiping procedure Injection of clean water Supply of external powerDescription of Each Part Preparation for OperationWarm-up time Setting of temperature inside lockerAirtightness test INDICATION/OPERATION Panel SO2Function and indication of indication/operation panel This panel consists of the following functionsAdvice on Operation General operation of indication/operation panelPreparation of standard gas CalibrationKinds of standard gas Pressure regulation for standard gas cylinder Calibration concentration setting except zirconia O2 sensor Func CAL SET ENTCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operating procedure OperationMaintenance Routine maintenance and inspectionConditioner Periodic maintenance and inspection Other maintenance and inspectionCentration is measured in the setting range of 0 to 200ppm or more at 0.5L/minMaintenance of sampling device Gas extractor type ZBAK1Membrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Maintenance of zirconia O2 sensor Others, major component figuresStandard Accessories and Spares Standard accessoriesSpares for 1 year optional Code for preparation of 1-year spares and suppliesProcessing of Heating Pipe Ends OutlineEnd processing method Grid 300mm 40mm Grid area 150 mm Check after piping Clamp supportSpecifications of tube trace Specifications of heating pipeTroubleshooting Troubleshooting of instrumentError codes and remedies Ambient temperature Power supplyPower consumption Mass weightNumber of output points and kinds of outputs Standard requirements For sample gasMonth/NO2, 100ppm Block diagram Moving average outputAutomatic calibration Ordering informationCalibration tolerance Automatic calibration output hold Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For outdoor installation type For indoor installation typeConnection Diagram Only specified outputs are wiredHead Office Instrumentation Div International Sales Dept