Fuji Labs ZSP General operation of indication/operation panel, Advice on Operation

Page 17

7.3General operation of indication/operation panel

The measuring mode is assumed when power is turned ON. The gas concentra- tion appears on the main indication, while the range being used appears on the sub in- dication.

NO

Gas concentration

Range

indication

indication

Advice on Operation

When selecting the range

In the setting status, as shown in the figure at the right, the high range is selected when key is pressed, while the low range is selected when key is pressed.

Hige rang is selected when key is pressed

Low rang is selected when key is pressed

When selecting the gas component

In the setting status, as shown in the figure

at the right, the gas component can be set

by pressing COMP key.

Example: When COMP key is pressed

while the 1st component is flash-

ing, the flashing moves to the

2nd component (figure at right).

The setting for the 2nd compo-

nent is now changeable.

* The 1 component type analyzer is not pro-

COMP

ENT ZERO SPAN CAL

1st component

2rd component

3rd component

vided with COMP key.

For releasing zero or span

If a mistake has been made in zero or span

calibration, then perform a reset in the fol-

lowing way.

The figure at the right shows an example.

If SPAN key has been pressed mistakenly

instead of ZERO key, then press SPAN

key again. Calibration is released.

If ZERO key has been pressed mistak-

enly instead of SPAN key, the press

ZERO key again.

Calibration is released.

Pressing changes the settable gas component.

ZERO SPAN

Pressed wrongly.

ZERO SPAN

The lamp blinks.

ZERO SPAN

Press again to extinguish the lamp.

Thus reset. Press the ZERO key afresh.

– 9 –

Image 17
Contents Infrared GAS Analyzer Request IntroductionProhibition Iii Prohibition Others Contents Vii Outline SpecificationsComposition PreparedSelection of gas sampling location InstallationSelection of installation site Connection of cables Installation workPiping procedure Supply of external power Injection of clean waterDescription of Each Part Preparation for OperationWarm-up time Setting of temperature inside lockerAirtightness test SO2 INDICATION/OPERATION PanelThis panel consists of the following functions Function and indication of indication/operation panelGeneral operation of indication/operation panel Advice on OperationPreparation of standard gas CalibrationKinds of standard gas Pressure regulation for standard gas cylinder Func CAL SET ENT Calibration concentration setting except zirconia O2 sensorCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operation Operating procedureRoutine maintenance and inspection MaintenanceConditioner Other maintenance and inspection Periodic maintenance and inspectionCentration is measured in the setting range of 0 to 200ppm or more at 0.5L/minGas extractor type ZBAK1 Maintenance of sampling deviceMembrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Others, major component figures Maintenance of zirconia O2 sensorStandard accessories Standard Accessories and SparesCode for preparation of 1-year spares and supplies Spares for 1 year optionalOutline Processing of Heating Pipe EndsEnd processing method Grid 300mm 40mm Grid area 150 mm Clamp support Check after pipingSpecifications of heating pipe Specifications of tube traceTroubleshooting of instrument TroubleshootingError codes and remedies Power supply Ambient temperaturePower consumption Mass weightNumber of output points and kinds of outputs Standard requirements For sample gasMonth/NO2, 100ppm Moving average output Block diagramOrdering information Automatic calibrationCalibration tolerance Automatic calibration output hold Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For indoor installation type For outdoor installation typeOnly specified outputs are wired Connection DiagramHead Office Instrumentation Div International Sales Dept